BIOMETRIA OTTICA IOL PREMIUM E NON SOLO · Nei bulbi miopi elevati Zonula debole+liquefazione...
Transcript of BIOMETRIA OTTICA IOL PREMIUM E NON SOLO · Nei bulbi miopi elevati Zonula debole+liquefazione...
Scipione Rossi Ospedale San Carlo, GVM care & Research
Roma
Quando le cose non vanno.
Gestione degli errori refrattivi
BIOMETRIA OTTICA IOL PREMIUM E NON SOLO
ELEMENTI CHIAVE PER IL SUCCESSO DELLE IOL PREMIUM
Obbiettivo: EMMETROPIA POST-OP spherical equivalent −0.5 to +0.5 D and <0.75 D astigmatism
Gatinel , 2016
SORPRESA REFRATTIVA DOPO IMPIANTO DI UNA PREMIUM – IOL SFERICA
Alió JL, Ahmed A, Abdelghany AA, Fernández-Buenaga R: Management of residual refractive error after cataract surgery. Curr Opin Ophthalmol 2014, 25(4):291–297.
Abdelghany A.A.: Alio J.L:Surgical options for correction of refractive error following cataract surgery .Eye and Vision, 2016
Denoyer A, Ricaud X, Van Went C, Labbe A, Baudouin C: Influence of corneal biomechanical properties on surgically induced astigmatism in cataract surgery. J Cataract Refract Surg
2013, 39(8):1204–1210.
Vries NE, Webers CA, Touwslager WR, et al. Dissatisfaction after implantation of multifocal intraocular lenses. J Cataract Refract Surg. 2011;37:859-865.
Woodward MA, Randleman JB, Stulting RD. Dissatisfaction after multifocal intraocular lens implantation. J Cataract Refract Surg. 2009;35:992-997.
Hayashi K, Manabe S, Yoshida M, Hayashi H. Effect of astigmatism on visual acuity in eyes with a diffractive multifocal intraocular lens. J Cataract Refract Surg. 2010;36:1323-
1329.
Ben Yaish S, Zlotnik A, Raveh I, Yehezkel O, Belkin M, Zalevsky Z. Intraocular omni-focal lens with increased tolerance to decentration and astigmatism. J Refract Surg. 2010;26:71-
76
DEFOCUS & ASTIGMATISMO POST-OP
• ERRORI DI CALCOLO
• FATTORI CHIRURGICI NON PREDITTIBILI
“the effective lens position, iol tilt, surgically induced astigmatism (SIA) ,
corneal hysteresis and biomechanical features also modulates optical
changes”
ANALISI DEI RISULTATI POST-OPERATORI
1. Refrazione Manifesta: MRx
2. Analisi aberrometrica interna (in midriasi)
3. Topo/tomografia
4. Posizione
centraggio/allineamento se T- IOL
INSUCCESSO DOPO IMPIANTO DI UNA IOL PREMIUM
RESIDUO SFERO-CILINDRICO
Cosa fare? … elemento chiave: Identificare la causa
Gundersen K.G., Makari S., Ostenstad S.:retreatments after multifocal intraocular lens implantation: an analysis Clinical
Ophthalmology 2016:10 365–371
• ABERRAZIONI DI BASSO ORDINE
legate ad errori di calcolo
• ABERRAZIONI DI ALTO ORDINE
legate al centraggio ed allineamento
Analisi del fronta d’onda interno
(esame in midriasi)
Reale performance della IOL
Elimina fattori pseudoaccomodativi
ANALISI DEL RISULTATO OTTENUTO CON PREMIUM-IOL
CHECK: centraggio e qualità della visione
INSUCCESSO DOPO IMPIANTO PREMIUM IOL SFERICA RESIDUO SFERO-CILINDRICO REFRATTIVO
3 OPZIONI CHIRURGICHE
1) CHIRURGIA REFRATTIVA CORNEALE
2) POLIPSEUDOFACHIA SECONDARIA (IOL PIGGY BACK)
3) IOL EXCHANGE
OBBIETTIVO: valutare la migliore strategia per correggerlo
Gundersen K.G., Makari S., Ostenstad S.: Retreatments after multifocal intraocular lens implantation: an analysis Clinical
Ophthalmology 2016:10 365–371
CHIRURGIA REFRATTIVA CORNEALE LASER
CORRETTO TIMING
DOPO 6 MESI DALLA CHIRURGIA DELLA CATARATTA
• STABILITÀ REFRATTIVA
• CORRETTO WOUNG HEALING DELLA MAIN
INCISION
Muftuoglu O, Prasher P, Chu C, et al. Laser in situ keratomileusis for residual refractive errors after apodized diffractive multifocal intraocular lens
implantation. J Cataract Refract Surg. 2009;35:1063– 1071.
Jendritza BB, Knorz MC, Morton S. Wavefront-guided excimer laser vision correction after multifocal IOL implantation. J Refract Surg.
2008;24:274–279.
PROFILO DI ABLAZIONE SEMPLICE
AMETROPIE DI LIEVE ENTITÀ (Refrazione Manifesta)
HOAS CORNEALI E INTERNE <0,5 MICRON
CHIRURGIA REFRATTIVA CORNEALE LASER
INDICAZIONI
ABLAZIONE laser STANDARD
ABLAZIONE WAVEFRONT GUIDATA IN BULBO PSEUDOFACHICO: PROBLEMATICHE
DIFFICOLTÀ DI ACQUISIZIONE DEL FRONTE
D’ONDA CON GLI ABERROMETRI HARTMANN
SHACK
POSSIBILE CROSSOVER DEL PATTERN DEI
CENTROIDI
ARTEFATTI CREATI DALLA IOL (OPACITÀ CP)
DAL MARGINE DELLA CAPSULOTOMIA SE
ASIMMETRICA E NON CENTRATA
Schallhorn S.C., Venter G.A., Teenan S.et al.: Outcomes of excimer laser enhancements in pseudophakic patients with multifocal
intraocular lens . Clinical Ophthalmolgy, 2016
CENTRAGGIO PERFETTO
CASO CLINICO RESIDUO IPERMETROPICO DOPO IOL TRIFOCALE
UCVA 4/10
BCVA 10/10
MRx+1,5 SF ANALISI ABERROMETRICA
SOLO DEFOCUS INTERNO SENZA HOAS
UCVA 10/10 dopo FS-LASIK standard
AT LISA
Potere +12
500micron c.a. =
+1.50
Nei bulbi miopi elevati
Zonula debole+liquefazione vitreo
VARIABILITÀ DELL’ELP
GUO T., GAO P., FANG L. ET AL.: EFFICACY OF TORIC INTRAOCULAR LENS IMPLANTATION IN EYES WITH HIGH MYOPIA: A PROSPECTIVE, CASE‐CONTROLLED
OBSERVATIONAL STUDY EXPERIMENTAL AND THERAPEUTIC MEDICINE 15: 5288-5294, 2018
Posizione effettiva
della IOL
Corretta posizione
della IOL
STIMA DELL’ELP CRITICA!
POLIPSEUDOFACHIA SECONDARIA
Piggy Back IOL
Jin H, Limberger IJ, Borkenstein AF, Ehmer A, Guo H, Auffarth GU. Pseudophakic eye with obliquely crossed piggyback toric intraocular lenses. J Cataract Refract
Surg. 2010;36(3):497- 502
CORREZIONE DI UN ELEVATO RESIDUO REFRATTIVO SFERICO POST-OPERATORIO
PIGGY BACK IOL VANTAGGI
Khan MI, Muhtaseb M. Performance of the sulcoflex piggyback intraocular lens in pseudophakic patients. J Refract Surg. 2011;27(9): 693–696.
Venter JA, Oberholster A, Schallhorn SC, Pelouskova M. Piggyback intraocular lens implantation to correct pseudophakic refractive error after segmental multifocal
intraocular lens implantation. J Refract Surg. 2014;30(4):234–239.
RIABILITAZIONE VISIVA RAPIDA E ALTISSIMA PREDITTIBILITÀ REFRATTIVA
IL CALCOLO DEL POTERE SI BASA SUL RESIDUO REFRATTIVO PSEUDOFACHICO!
PER LA MIOPIA: EQUIVALENTE SFERICO X 1.15
PER L’IPERMETROPIA: EQUIVALENTE SFERICO X 1.25
Opacizzazione intralenticolare (ILO)
Jin H, Limberger IJ, Borkenstein AF, Ehmer A, Guo H, Auffarth GU. Pseudophakic eye with obliquely crossed piggyback toric intraocular lenses. J Cataract Refract
Surg. 2010;36(3):497- 502
PIGGY BACK IOL DI VECCHIA GENERAZIONE PROBLEMATICHE
Diametro dell’ottica 6.5mm, lunghezza 14 mm
Ridotto rischio di cattura dell’ottica
Alta stabilità rotazionale
Aptiche ondulate angolate 10°
Ridotto rischio contatto uveale
Ridotto rischio S. da dispersione pigmentaria
ottica monofocale, torica e multifocale
Yuen L., Jennifer Hung j.:supplementary toric intraocular lens implantation for pseudophakic refractive error
JCRS Online Case Reports 2015; 3:68–73
SUPERFICIE POSTERIORE DELL’OTTICA CONCAVA
PIGGY BACK IOL DI NUOVA GENERAZIONE
3) IOL EXCHANGE
INDICAZIONI:
1) ABERRAZIONI BASSO ORDINE
DEFOCUS E ASTIGMATISMO
ipo-ipercorrezioni sf o cyl
2) ABERRAZIONI ALTO ORDINE
COMA, CYL SECONDARI, TREFOIL
aloni, glare, disfotopsie…
Decentramento, disallineamento, tilt iol…
IOL EXCHANGE COMPLICANZE
• SCARSA PREDITTIBILITÀ DELL’ELP DEL NUOVO IMPIANTO
spostamento del diametro equatoriale del sacco
• ROTTURA SACCO CAPSULARE
• DEISCENZA ZONULARE
• PERDITA DI VITREO
• EDEMA MACULARE CISTOIDE
Aliò JL, Abdelghany AA, Fernandez-Buenaga R. Management of residual refractive error after cataract surgery. Curr Opin
Ophthalmol 2014; 25:291–297
Page T.P: Intraoperative aberrometry-assisted toric IOL exchange following unexpected surgically induced astigmatism JCRS Online
Case Reports 2014
CON IOL TRIFOCALI
TARGET POST OP.: PLANO
CON EDOF IOLS
• TOLLERANZA ALL’ERRORE REFRATTIVO (+/- 0,75 SF)
evitare ipermetropia (distanza lettura per vicino troppo elevata)
no miopia (distanza lettura per vicino troppo ridotta)
• TARGET OCCHIO DOMINANTE PLANO, TARGET OCCHIO PER VICINO -0,5
(MICROMONOVISIONE)
TARGET PREFRATTIVO POST-OP CON PREMIUM IOL
SCONTENTA PER RISULTATO VISIVO A
TUTTE LE DISTANZE
OO UCVA 4/10 OO BCVA 10/10 -0,75
ESEGUITO TEST CON LAC
STRATEGIA DI PRIMA SCELTA: LIFT DEL
FLAP + CORREZIONE LASER STANDARD
STRATEGIA DI SECONDA SCELTA: PIGGY
BACK
CASO CLINICO: DONNA 45 AA residuo miopico post pseudofachia trifocale (2018)+
capsulotomia posteriore in pregressa lasik miopica (2009)
1°mese post op 3° mese post op
Circa il 10% ha un astigmatismo residuo >1 D*
70% ha un residuo ≤ 0,75**
ASTIGMATISMO RESIDUO DOPO IMPIANTO DI UNA PREMIUM IOL TORICA
* Thomas BC, Khoramnia R, Auffarth GU, Holzer MP. Clinical outcomes after implantation of a toric intraocular lens with a transitional conic toric surface. Br J Ophthalmol
2017
**Holland E., LaneS., Horn J.D. et al.: the acrysof toric iol in subjects wiyh cataract and corneal astigmatism: a randomized subjectmasked parallel-group , 1 year study.
Ophthalmolgy, 2014;117:2104-2111
**Visser N, Bauer NJ, Nuijts RM. Toric intraocular lenses: historical overview, patient selection, IOL calculation, surgical techniques, clinical outcomes, and complications. J
Cataract Refract Surg. 2013; 39(4):624–637.
ANALISI DEL RISULTATO OTTENUTO CON PREMIUM IOL TORICHE
CHECK: centraggio e asse di allineamento della T-IOL in midriasi
1. BIOMICROSCOPIA
2. ABERROMETRIA INTERNA
ANALISI DELL’ALLINEAMENTO
ANALISI DEL CENTRAGGIO
CYL CORNEA
CYL T-IOL CYL TOT
OFFSET DELLA T-IOL DALL’ASSE VISIVO Presenza di HOAS INTERNE
Riduzione qualità e quantità della acuità visiva
POSIZIONE DELLA T-IOL
CORRETTA
ERRORE
CALCOLO SFERICO
ERRORE
CALCOLO CILINDRO
ERRATA
T-IOL
RUOTATA
T-IOL
DECENTRATATA
T-IOL
TILTATA
IPER-IPO CORREZIONE
Sferoequivalente >o< al target programmato Sferoequivalente = al target programmato
ALGORITMO DI ANALISI DEL RESIDUO SFERO CILINDRICO
CHIRURGIA REFRATTIVA CORNEALE LASER LIMITI
PROFILO D’ABLAZIONE COMPLESSO : ANOMALIE SFEROCILINDRICHE ELEVATE
dopo impianto iol multifocali toriche
1. L’ASTIGMATISMO REFRATTIVO IL CUI ASSE È LA SOMMA VETTORIALE TRA ASTIGMATISMO CORNEALE E
L’ASTIGMATISMO INTERNO LOAS Z(2±2)
2. LE ABERRAZIONI DI ALTO ORDINE INTERNE (HOAS)
ABLAZIONE CUSTOMIZZATA
WAVEFRONT GUIDATA
ABLAZIONE WAVEFRONT GUIDATA IN BULBO PSEUDOFACHICO: PROBLEMATICHE
DIFFICOLTÀ DI ACQUISIZIONE DEL FRONTE
D’ONDA CON GLI ABERROMETRI HARTMANN
SHACK
POSSIBILE CROSSOVER DEL PATTERN DEI
CENTROIDI
ARTEFATTI CREATI DALLA IOL (OPACITÀ CP)
DAL MARGINE DELLA CAPSULOTOMIA SE
ASIMMETRICA E NON CENTRATA
Schallhorn S.C., Venter G.A., Teenan S.et al.: Outcomes of excimer laser enhancements in pseudophakic patients with multifocal
intraocular lens . Clinical Ophthalmolgy, 2016
UN OCCHIO PSEUDOFACHICO CON UNA T-IOL RUOTATA È UN SISTEMA LENTE SOTTILE
COMPOSTO DI 2 SFEROCILINDRI OBLIQUAMENTE CROCIATI
IL RISULTATO È UN NUOVO SFEROCILINDRO
POTERE ASTIGMATICO PUÒ ESSERE COMPENSATO
DA UN 3° SFEROCILINDRO
Holladay JT, Cravy TV, Koch DD. Calculating the surgically induced refractive change following ocular surgery. J Cataract Refract Surg 1992; 18:429–443
Jin H., Limberger I.J , Ehmer A., Guo H., Auffarth G.U.:Impact of axis misalignment of toric intraocular lenses on refractive outcomes after cataract surgeryJ
Cataract Refract Surg 2010; 36:2061–2072
Holladay- Cravy-Koch mathematical method
PIGGY BACK CON SULCOFLEX TORICA
2AIOL TORICA QUALE POTERE? QUALE ASSE?
The Rayner online calculator
FOGLIO DI CALCOLO ON LINE SULCUFLEX TORICA
POTERE CALCOLATO DALLA REFRAZIONE MANIFESTA
ASSE 90° RISPETTO AL CILINDRO REFRATTIVO
RESIDUO SFEROCILINDRICO REFRATTIVO DOPO IMPIANTO FLACS CON SYMFONY TORICA
PRE-OP MRX-3sf-2 (50)
POST-OP:
UCVA 3/10
BCVA 7/10
MRx+1,25sf -2,50 cil (140)
25/11/18 18:54Calcolatore Express per lenti toriche
Pagina 1 di 1https://www.amoeasy.com/toric2(bD1pdCZjPTA1MA==)/ToricPrintResu…lDU1lNRk8mcF9pb2x0b29yZGVyPTAwMSZwX2NvdW50cnljb2RlPVpFTQ%3d%3d
V:1.51
Tornare al calcolatore V:1.51
Nome chirurgo Informazioni paziente Data Età paziente
rossi di paola 25.11.2018 22
Dettagli relativi alle IOL
Rifrazione residuaModello IOL Orientamento Cilindro Asse
ZXT225 142 ° -0,48 D 52°
Dattagli di calcoloPotere SE di IOL Indice K Convenzione per cilindro refrattivo
+20,00 1,3375 Meno
Cheratometria e biometria@ Asse piatto
(K1)
@ Asse curvo
(K2)SIA @ Asse
Lunghezza
assiale
Metodo
biometrico
Costante
A
Astigmatismo
corneale posteriore
45,18@45° 46,81@135° 0,25@30° 24,22Optical orImmersion
119,30 Sì
Dettagli relativi alle IOL Astigmatismo residuoModello IOL Orientamento Cilindro Asse
ZXT225 142 ° -0,48 D 52 °
ZXT300 142 ° -0,03 D 142 °
ZXT375 142 ° -0,53 D 142 °
Note
OS (Sinistro)
Temporal
Nasal
(sede dell’incisione) 30°
Orientamento calcolato
142 °
DISALLINEAMENTO T-IOL di quanti gradi?
WAVEFRONT CYL INT
T-IOL
WAVEFRONT CYL CORNEA
OFF AXIS 45°
cosa fare?
VALUTAZIONE ABERROMETRICA POST-OP ESAME IN MIDRIASI
149°
RIPOSIZIONAMENTO T-IOL
IL TIMING DELLA CHIRURGIA È CRUCIALE
DOPO 2-4 SETTIMANE ADESIONE TRA APTICHE E SACCO
3 ° MESE POST-OP SINDROME CONTRAZIONE SACCO CAPSULARE
Lockwood G.C., Randlema J.B.: Toric intraocular lens rotation to optimize refractive outcome despite appropriate intraoperative positioning J.
Cataract Refract Surg 2015; 41:878–883
Oshika T., Inamura n., Inoue Y. Et al.:Incidence and Outcomes of Repositioning Surgery to Correct Misalignment of Toric Intraocular Lenses.
Ophthalmology. 2018 Jan;125(1):31-35.
RIPOSIZIONAMENTO CON Z-ALIGN DOPO 7 GG
DOPO RIPOSIZIONAMENTO UCVA 10/10
25/11/18 18:54Calcolatore Express per lenti toriche
Pagina 1 di 1https://w
ww.am
oeasy.com/toric2(bD1pdC
ZjPTA1MA==)/ToricPrintResu…
lDU1lNRk8m
cF9pb2x0b29yZGVyPTAw
MSZw
X2NvdW
50cnljb2RlPVpFTQ%
3d%3d
V:1.51
Tornare al calcolatoreV:1.51
Nome chirurgo
Informazioni paziente
DataEtà paziente
rossidi paola
25.11.201822
Dettagli relativi alle IOL
Rifrazione residua
Modello IO
LO
rientamento
CilindroAsse
ZXT225142 °
-0,48 D52°
Dattagli di calcoloPotere SE di IO
LIndice K
Convenzione per cilindro refrattivo
+20,001,3375
Meno
Cheratometria e biom
etria@
Asse piatto
(K1)
@ Asse curvo
(K2)SIA @
AsseLunghezza
assiale
Metodo
biometrico
Costante
A
Astigmatism
o
corneale posteriore
45,18@45°46,81@
135°0,25@30°
24,22O
ptical orIm
mersion
119,30Sì
Dettagli relativi alle IOL
Astigmatism
o residuoM
odello IOL
Orientam
entoCilindro
Asse
ZXT225142 °
-0,48 D52 °
ZXT300142 °
-0,03 D142 °
ZXT375142 °
-0,53 D142 °
Note
OS (Sinistro)
Temporal
Nasal
(sede dell’incisione) 30°
Orientam
ento calcolato
142 °
CALCOLATORI DI ANALISI VETTORIALE ON LINE
WWW.ASTIGMATISMFIX.COM (BERDAHL AND HARDTEN)
WWW.ASSORT.COM
BARRETT RX FORMULA
OTTIMIZZAZIONE DEL RIPOSIZIONAMENTO DI UNA T-IOL
Alpins N, Ong JKY, Stamatelatos G. Refractive surprise after oric intraocular lens implantation: graph analysis. J Cataract Refract
Surg 2014; 40:283–294
ABERROMETRIA INTRAOPERATORIA
Optiwave Refractive Analysis (ORA)
system
33
www.astigmatismfix.com
www.astigmatismfix.com
CONSIDERAZIONI
ERRORI BIOMETRICI, ERRORI DI POSIZIONAMENTO , ERRORI CHIRURGICI OPPURE
ROTAZIONE SPONTANEA DI UNA T-IOL POSSONO DETERMINARE RESIDUI
REFRATTIVI SFREOCILINDRICI
IN CASO DI RESIDUO POST-OPERATORIO E SEMPRE OPPORTUNO RIPERCORRERE
TUTTE LE PROCEDURE DI CALCOLO DELLA IOL
CONSIDERAZIONI
LA REFRAZIONE E L’OSSERVAZIONE DELL’ALLINEAMNETO DELL’ASSE DEVE ESSERE
RICERCATA GIÀ LA PRIMA ORA DOPO L’INTERVENTO
SE IL RIALLINEAMENTO È RICHIESTO (SFERO EQUIVALENTE=0) DOVREBBE ESSERE
FATTO ENTRO 4 SETTIMANE DALL’IMPIANTO
IN CASO DI ADESIONE APTICHE/SACCO LA CREAZIONE DI CHERATOTOMIE
ARCUATE FS LASER ASSISTITE È UN’OPAZIONE EFFICACE E SICURA
CONSIDERAZIONI
IN CASO DI ERRORI DI CALCOLO LA CORREZIONE DEL RESIDUO SFEROCILINDRICO IN
UN BULBO PSEUDOFACHICO PUÒ ESSERE EFFETTUATA CON UN NUOVO
SFEROCILINDRO
ESEGUI UN’ABLAZIONE WAVEFRONT GUIDATA NEI CASI IN CUI L’ACQUISIZIONE
ABERROMETRICA SIA POSSIBILE
OPPURE ESEGUI L’ IMPIANTO DI UNA SECONDA IOL : TECNICA ADDITIVA,
REVERSIBILE E SICURA
CONSIDERAZIONI
LA IOL EXCHANGE RAPPRESENTA LA SOLUZIONE FINALE AL PROBLEMA
RIPETERE LA BIOMETRIA E CONFRONTARE I RISULTATI
NELLA SCELTA DEL NUOVO POTERE DELLA IOL TENERE CONTO DELLA
REFRAZIONE RESIDUA E VALUTARE LA CONCORDANZA DEI DATI (1 D SULLA
IOL= 0.75 D REFRAZIONE) (FORMULE DI VERGENZA)
CONCLUSIONI
PER OTTENERE UN’ALTA PREDITTIBILITÀ DEI RISULTATI DOPO IMPIANTO DI T-IOL
1. BIOMETRIA IMPECCABILE (FORMULECHE STIMANOL’ELP)
2. USA FOGLI DI CALCOLO DI ULTIMA GENERAZIONE (SUPERFICIE POSTERIORE
CORNEALE)
3. ESEGUI UN’ACCURATA CHIRURGIA SE POSSIBILE FS LASER GUIDATA
4. EFFETTUA UN ALLINEAMENTO DIGITALE INTRAOPERATORIO DELLA T-IOL
5. SCEGLI UN T-IOL CON UNA GRANDE STABILITÀ ROTAZIONALE……
GRAZIE PER L’ATTENZIONE