1 P.O. dei Pellegrini; A.S.L. Napoli 1Centro 2 Dipartimento di Oftalmologia, Seconda Università degli Studi Napoli Scheimpflug topography , Scanning-Slit topography and Anterior Segment Oct in patients with keratoconus Luigi Capasso, MD 1 ; M Lanza, MD 2 ; A Tortori, MD 1 ; C Buono, MD 1 ;L Gifuni, MD 1 ;F Amoroso 1 ; N Rosa, MD 2 Authors have no financial interest.

description

Scheimpflug topography , Scanning-Slit topography and Anterior Segment Oct in patients with keratoconus. Luigi Capasso, MD 1 ; M Lanza, MD 2 ; A Tortori, MD 1 ; C Buono, MD 1 ;L Gifuni, MD 1 ;F Amoroso 1 ; N Rosa, MD 2. 1 P.O. dei Pellegrini; A.S.L. Napoli 1Centro - PowerPoint PPT Presentation

Transcript of 1 P.O. dei Pellegrini; A.S.L. Napoli 1Centro

Page 1: 1  P.O. dei Pellegrini; A.S.L. Napoli 1Centro

1 P.O. dei Pellegrini; A.S.L. Napoli 1Centro2 Dipartimento di Oftalmologia, Seconda Università degli Studi Napoli

Scheimpflug topography , Scanning-Slit topography and

Anterior Segment Oct in patients with keratoconus

Luigi Capasso, MD1; M Lanza, MD2 ; A Tortori, MD1; C Buono, MD1;L Gifuni, MD1;F Amoroso1; N Rosa, MD2

Authors have no financial interest.

Page 2: 1  P.O. dei Pellegrini; A.S.L. Napoli 1Centro

BACKGROUND

Accurate measurement of corneal thickness

is crucial •in the monitoring ectatic dystrophies •in the preoperative assessment•in the planning of surgical procedures such as intrastromal corneal ring segment implantation, corneal collagen cross-linking with riboflavin (C3-R) and deep lamellar keratoplasty.

Page 3: 1  P.O. dei Pellegrini; A.S.L. Napoli 1Centro

PURPOSE

To compare the measurements of the thinnest corneal

point (TCP) with three different devices

Orbscan II Computerised Topography, Visante Optical

Coherence Tomography (OCT) and Pentacam

in patients with keratoconus (KC)

Page 4: 1  P.O. dei Pellegrini; A.S.L. Napoli 1Centro

prospective, clinical study, 71 eyes of 37 consecutive patients with keratoconus.

MATERIAL AND METHODS

25 males and 12 females

Keratoconus was diagnosed if there was scissoring reflex on retinoscopy and central or paracentral steepening of the cornea on computerized topography with at least one of the following slit lamp findings of keratoconus: central or paracentral thinning, anterior bulging or conicity, Fleisher’s ring. No patient presented Descemet’s breaks or apical scar and subepithelial fibrosis.We studied a spectrum of keratoconic eyes of different severity according to the Amsler scale automatically provided by the Pentacam as follows: early (Amsler 1) 23 eyes; moderate (Amsler 2) 26 eyes; and severe (Amsler 3-4) 22 eyes.

Age range: 12 – 29 years (mean= 22.01 ± 4.53) years

Page 5: 1  P.O. dei Pellegrini; A.S.L. Napoli 1Centro

MATERIAL AND METHODS

Measurements were done sequentially by two different operators using Orbscan II computerised topography (Bausch and Lomb Inc), Visante anterior segment OCT (Carl Zeiss) and Pentacam (Oculus).

The software of the three devices automatically generates a pachimetry map.

All measurements were taken at the same day time, between 1 PM and 5 PM.

Page 6: 1  P.O. dei Pellegrini; A.S.L. Napoli 1Centro

RESULTSThe TCP measured with the Orbscan II and Visante OCT was 424.74 ± 54.91 μm (mean ± SD) and 461 ± 38.48 μm respectively, with a statistically significant difference (p<0.001).

Page 7: 1  P.O. dei Pellegrini; A.S.L. Napoli 1Centro

RESULTSThe TCP measured with the Pentacam and Visante OCT was 461.94 ± 38.94 μm (mean ± SD) and 461 ± 38.48 μm respectively, with no statistically significant difference (p>0.1).

Page 8: 1  P.O. dei Pellegrini; A.S.L. Napoli 1Centro

RESULTSThe TCP measured with the Orbscan II and Pentacam was 424.74 ± 54.91 μm (mean ± SD) and 461.94 ± 38.94 μm respectively, with a statistically significant difference (p<0.001).

Page 9: 1  P.O. dei Pellegrini; A.S.L. Napoli 1Centro

Discrepancies between the three devices were present, but Pentacam and Visante OCT, showing no significant differences, seem to be more reliable in measuring the corneal thinnest point in patients with keratoconus compared to Orbscan II.

CONCLUSIONS