Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione)...

9
www. .com 68 Data Logger Multicanale Modello LR8410/20 LR8400/20 LR8401/20 LR8402/20 LR8431/20 8423 Misure Tensione Vcc Portate: ±10mV a ±100V Portate: ±10mV a ±100V Portate: ±10mV a ±100V Portate: ±10mV a ±100V Temperatura termocoppie K, J, E, T, N, R, S, B, da -200°C a +2000°C K, J, E, T, N, R, S, B, W, da -200°C a +2000°C K, J, E, T, N, R, S, B, da -200°C a +2000°C K, J, E, T, N, R, S, B, da -200°C a +2000°C Temperatura termoresistenze PT100 e jPT100, da -200°C a +800°C PT100 e jPT100, da -200°C a +800°C - PT100 e jPT100, da -200°C a +800°C Umidità con sensore 2000Z da 0% a 100% U.R. con sensore 2000Z da 0% a 100% U.R. - con sensore 9701 da 0% a 100% U.R. Resistenza Rdc Portate: da 10a 200Portate: da 10a 200- - Impulsi - 8 canali 8 canali 8 canali 4 canali 120* Ingressi logici - 8 canali 8 canali 8 canali - 120* Prestazioni di misura e registrazione Velocità di campionamento da 10msec a 60 min da 10msec a 60 min da 10msec a 60 min da 10msec a 60 min Memoria interna 16MB 16MB 16MB 16MB 7MB 32MB Card 2GB 2GB 2GB 2GB 2GB 1GB Ingressi di misura Ingressi isolati tra loro SI* SI* SI* SI* SI SI* Tensione max tra canali 300Vcc 300Vcc* 300Vcc* 300Vcc* 60Vcc 200Vcc* Tensione max verso terra 300Vcc/ca 300Vcc/ca 300Vcc/ca 300Vcc/ca 60Vcc 600Vcc/ca* Max ingressi analogici 105 60 60 60 10 120 Max ingressi digitali - 8 8 8 4 (solo impulsi) 120 Moduli di ingresso Max 7, con Bluetooth Max 4 per totale 60 canali analogici - Max 8 da 15 canali Display Dimensioni display grafico 5.7 pollici 5.7 pollici 5.7 pollici 5.7 pollici 4.3 pollici su PC tramite software Interfacce USB SI SI SI SI SI SI Slot per chiavi USB SI SI SI SI SI - LAN SI SI SI SI SI SI SD Card SI - - - - - CF Card - SI SI SI SI SI Alimentazione Diretta in CA - - - - - SI Tramite adattatore in CA SI SI SI SI SI SI Tramite batterie ricaricabili** SI SI SI SI SI - Diretta in CC SI SI SI SI SI - (*) le caratteristiche indicate con asterisco (*) sono da valutare in funzione dei moduli di ingresso intercambiabili (opzionali) installati sull’unità principale (**) non fornite in dotazione

Transcript of Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione)...

Page 1: Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione) Display 40 x 25 mm Connettività a Data-Logger tramite Bluetooth® a LR8410/20 Alimentazione

www. .com68

Data Logger Multicanale

Modello LR8410/20 LR8400/20 LR8401/20 LR8402/20 LR8431/20 8423

Misure

Tensione Vcc Portate: ±10mV a ±100V

Portate: ±10mV a ±100V

Portate: ±10mV a ±100V

Portate: ±10mV a ±100V

Temperatura termocoppieK, J, E, T, N, R, S, B,

da -200°C a +2000°C

K, J, E, T, N, R, S, B, W, da -200°C a +2000°CK, J, E, T, N, R, S, B,

da -200°C a +2000°C

K, J, E, T, N, R, S, B,da -200°C a +2000°C

Temperatura termoresistenze PT100 e jPT100,da -200°C a +800°C PT100 e jPT100, da -200°C a +800°C - PT100 e jPT100,

da -200°C a +800°C

Umidità con sensore 2000Zda 0% a 100% U.R. con sensore 2000Z da 0% a 100% U.R. - con sensore 9701

da 0% a 100% U.R.

Resistenza Rdc Portate:da 10Ω a 200Ω Portate: da 10Ω a 200Ω - -

Impulsi - 8 canali 8 canali 8 canali 4 canali 120*

Ingressi logici - 8 canali 8 canali 8 canali - 120*

Prestazioni di misura e registrazione

Velocità di campionamento da 10msec a 60 min da 10msec a 60 min da 10msec a 60 min da 10msec a 60 min

Memoria interna 16MB 16MB 16MB 16MB 7MB 32MB

Card 2GB 2GB 2GB 2GB 2GB 1GB

Ingressi di misura

Ingressi isolati tra loro SI* SI* SI* SI* SI SI*

Tensione max tra canali 300Vcc 300Vcc* 300Vcc* 300Vcc* 60Vcc 200Vcc*

Tensione max verso terra 300Vcc/ca 300Vcc/ca 300Vcc/ca 300Vcc/ca 60Vcc 600Vcc/ca*

Max ingressi analogici 105 60 60 60 10 120

Max ingressi digitali - 8 8 8 4 (solo impulsi) 120

Moduli di ingresso Max 7, con Bluetooth Max 4 per totale 60 canali analogici - Max 8

da 15 canaliDisplay

Dimensioni display grafico 5.7 pollici 5.7 pollici 5.7 pollici 5.7 pollici 4.3 pollici su PC tramite software

Interfacce

USB SI SI SI SI SI SI

Slot per chiavi USB SI SI SI SI SI -

LAN SI SI SI SI SI SI

SD Card SI - - - - -

CF Card - SI SI SI SI SI

Alimentazione

Diretta in CA - - - - - SI

Tramite adattatore in CA SI SI SI SI SI SI

Tramite batterie ricaricabili** SI SI SI SI SI -

Diretta in CC SI SI SI SI SI -

(*) le caratteristiche indicate con asterisco (*) sono da valutare in funzione dei moduli di ingresso intercambiabili (opzionali) installati sull’unità principale(**) non fornite in dotazione

Page 2: Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione) Display 40 x 25 mm Connettività a Data-Logger tramite Bluetooth® a LR8410/20 Alimentazione

www. .com 69

Mini Data Logger Bluetooth®

Modello LR8512 LR8513 LR8514 LR8515 LR8520

Misure

Tensione Vcc - - - ±50Vcc -

Corrente Acc - fino a 2000Acc - - -

Corrente Aca - fino a 1000Aca - - -

Temperatura - - -40°C … +80°C - -40°C … +80°C

Termoresistenze (K e T) - - - -200°C … +1000°C -

Umidità - - 0% … 100% - 0% … 100%

Conta-Impulsi/ Contagiri 2 canali - - - -

Indice fungino - - - - SI

Prestazioni di misura e registrazione

Quantità di canali 2 canali 2 canali 2+2 canali 2 canali 1+1 canali

Memoria interna 500.000 dati/canale 500.000 dati

Cadenza di registrazione da 0.1sec a 60min da 0.5sec a 60min da 0.1sec a 60min da 0.5sec a 60min

Tipo di registrazione Valore istantaneo Istantaneo e medio Valore istantaneo

Comunicazione e interfaccia

Tipo di connessione Bluetooth®2.1 + EDR

Dispositivi supportati Windows PC e Android tablet + smartphone

Sistemi Operativi Windows 8.1/8/7/Vista (32-64bit) – Android OS 4.0.3 o superiore

Software di analisi dati Logger Utility (in dotazione)

Display 40 x 25 mm

Connettività a Data-Logger tramite Bluetooth® a LR8410/20

Alimentazione

Tramite batterie Nr. 02 batterie alkaline LR6 (in dotazione)

Tramite adattatore in CA Alimentatore in CA (opzionale)

Esterna in CC Da 5Vcc a 13.5Vcc (anche tramite USB con apposito cavetto, non fornito)

Accessori in dotazione

Batterie LR06 02 02 02 02 02

Cavetteria L1010 (02) - - - L1010 (01)

Accessori opzionali

Sensori - 7 modelli, da 500mA a 2000A

Z2010Z2011 - Z2010

Z2011

Alimentatore in CA Z2003 (da 100Vca a 240Vca, 50-60Hz – uscita 12Vcc)

Supporto magnetico Z5004 (cinghia di fissaggio con supporto magnetico)

(*) le caratteristiche indicate con asterisco (*) sono da valutare in funzione dei moduli di ingresso intercambiabili (opzionali) installati sull’unità principale(**) non fornite in dotazione

Dat

a Lo

gger

Mul

tica

nale

Min

i Dat

a Lo

gger

Blu

etoo

th®

Page 3: Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione) Display 40 x 25 mm Connettività a Data-Logger tramite Bluetooth® a LR8410/20 Alimentazione

www. .com 79

LR8512: conta-impulsi conta-giri a 2 canali

Ideale per applicazioni ove è necessario controllare flussi e velocità di liquidi e fluidi quali acqua, gas ed oli, rotazioni di alberi motore, encoder e sensori di processo, nei settori industriali, termo-idraulici ed automotive.

LR8513: registratore di corrente CA/CC a 2 canali

Da abbinare a specifico sensore di corrente, può essere utilizzato per monitorare assorbimenti di linee elettriche, carrelli elevatori ed ascensori, linee ferroviarie, veicoli elettrici ed ibridi, dispersioni di corrente, ecc...

LR8514: temperatura + umidità a 4 canali (2+2)

L’utilizzo tipico di LR8514 è il monitoraggio continuativo delle condizioni ambientali di luoghi di stoccaggio, laboratori, magazzini ad ambiente controllato (stagionatura, essicatura, conservazione), celle frigorifere.

LR8515: tensione + temperatura a 2 canali

Ideale per applicazioni ove la simultaneità di controllo tra tensione e temperatura è fondamentale: impianti fotovoltaici, propulsori elettrici ed ibridi, ricarica di batterie.

LR8520: indice fungino in ambienti critici

Il mini data-logger LR8520, sfruttando le misure simultanee di temperatura e umidità, calcola il tasso di contaminazione fungina ed elabora il fattore di crescita delle muffe. Trova impiego tipico nello stoccaggio di alimenti, cereali, documenti, così come nelle aree espositive delicate quali gallerie d’arte e musei.

LR85

XX

NEWLR85xxMini-Logger Bluetooth®,

con gestione su APP Android

Page 4: Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione) Display 40 x 25 mm Connettività a Data-Logger tramite Bluetooth® a LR8410/20 Alimentazione

www. .com80

Dispositivi Android

Tablet, Smartphone

Massima praticitàL’interfaccia utente è ottimizzata per lavisualizzazione su displaydel dispositivo Android

Analisi dati sul postoI dati registrati possono essereanalizzati sul posto direttamentesul dispositivo Android

Impostazioni

Analisi dei trendTramite la APP è possibilevisualizzare i trend graficidurante la fase diregistrazione. Questa funzione è utile per controlli preliminari

Specifiche tecniche

Procedura

Supported devices Android tablet / Android smartphoneCommunications Bluetooth®2.1 + EDRAndroid OS 4.0.3 or later

Number of availableregistrations

Max. 100 units

Recommended display size

7 inches or larger

SoftwareCollection: Wireless Logger Collector for Android Analysis: Logger Utility (PC)

Software acquisitionCollection: Download from Google Play Analysis: Supplied CD-R / Download from HIOKI's website

Imposta e avviaDefinisci le condizioni di misura e registrazione ed inviale al logger per avviare il salvataggio

Salvataggio datiÈ possibile salvare i dati sul dispositivo Android mentre la registrazione è in corso

Analisi dei datiTrasferisci i file a PCe analizzali con “Logger Utility” software

2

3

1

USB

Windows PC

Computer

Gestione multi-logger“Logger Utility” èin grado di gestirefino a 100 logger,creando gruppi esotto-gruppi ordinaticon struttura ad albero

Registrazioneperiodica su PCLa cadenza di trasferimento a PCè configurabile da10 minuti a 1 giorno

Verifica della operativitàIn ogni momento puoi controllarela registrazione, la batteria residua e il segnale wireless

Specifiche tecniche

Supported devices Windows PC / Windows tabletCommunications Bluetooth®2.1 + EDROS Windows 8 / 7 / Vista (32/64bit)

Number of availableregistrations

Max. 100 units

SoftwareCollection: Wireless Logger CollectorAnalysis: Logger Utility

Software acquisition Supplied CD-R / Download from HIOKI’s website

ProceduraImposta e avviaDefinisci le condizioni di misura e registrazioneed inviale al logger per avviare il salvataggio

Salvataggio datiÈ possibile salvare i dati su PC Windows mentre la registrazioneè in corso

Analisi dei datiAvvia il software“Logger Utility” ed esegui le analisi

2

3

1

Page 5: Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione) Display 40 x 25 mm Connettività a Data-Logger tramite Bluetooth® a LR8410/20 Alimentazione

www. .com 81

Progettazione volta al risparmio di energia

Connettività wireless

30 metri in linea diretta, fino a 100 logger

La APP Android e il software“Logger utility” possono gestirefino a 100 logger, tramiteconnessione Bluetooth®.La distanza di connettività dipendeanche dal tablet o PC utilizzato.

Tutte le registrazioni su un unico file

Indipendentemente da quando vengono svolte le registrazioni, i dati vengono automaticamente combinati su un unico file così da avere immediatamente una reale visione dei trend.

1st time 2nd time 3rd time 4th time

Massima praticità, senza fili dedicati alla trasmissione dei dati

L’installazione di un data-logger all’interno di un quadro elettrico non è mai stata così semplice. Con la connessione Bluetooth® è possibile chiudere la portella e lasciare tutto in perfetto ordine.LR85xx sono ideali anche per utilizzo su apparati mobili e carrellati.

Tutti i dati su unico file

30m

Collection Collection Collection Collection

Bassi consumi che allungano la vita delle batterie

Attivando la connessione Bluetooth® solo per piccoli periodi di tempo, si aumenta la durata delle batterie.

Vita della batteria, con operatività continuativa LR8512 LR8513 LR8514 LR8515 LR8520

Registrazione ad 1 minuto, bluetooh ® OFF 2 mesi 3 mesi 3.5 mesi 2.5 mesi 3.5 mesi

Registrazione ad 1 secondo, bluetooh ® OFF 2 mesi 1 mese 3 mesi 10 giorni 3 mesi

Registrazione ad 1 minuto, bluetooh ® ON 14 giorni 10 giorni 20 giorni 7 giorni 20 giorni

Con Bluetooth® sempre ON o sempre OFF. Per registrazioni prolungate, si consiglia l’utilizzo dell’alimentatore mod. Z2003.

Compatti, ad elevata capacità di memorizzazione

Posizionamento in spazi ristretti

Il formato tascabile e la cinghiamagnetica consentono ilposizionamento in luoghinascosti e poco fruibili.

500.000 dati per ogni canale di misura Doppia modalità di registrazione

I mini data-logger LR85xx offrono elevata capacità di memorizzazione per registrazioni approfondite, dettagliate, e di lunga durata.

Registrazione one-time: quando la memoria interna è piena, la registrazione si arresta.Registrazione continuativa: quando la memoria interna è satura, i nuovo dati vengono sovrascritti su quelli più lontani (FIFO). La capacità di memorizzazione è fissa a 500.000 valori/canale.Recording intervals Recordable time

0.1 sec 13 hr, 53 min, 20 sec

0.2 sec 1 day, 3 hr, 46 min, 40 sec

0.5 sec 2 days, 21 hr, 26 min, 40 sec

1 sec 5 days, 18 hr, 53 min, 20 sec

2 sec 11 days, 13 hr, 46 min, 40 sec

5 sec 28 days, 22 hr, 26 min, 40 sec

10 sec 57 days, 20 hr, 53 min, 20 sec

20 sec 115 days, 17 hr, 46 min, 40 sec

30 sec 173 days, 14 hr, 40 min, 00 sec

1 min 347 days, 5 hr, 20 min, 00 sec

2 min to 60 min Over 365 days

Temperatura Voltaggio

Tutti i modelli incorporano almeno 2 canali

Ogni modello di mini-loggerLR85xx dispone di 2 canali di ingresso per la misura e registrazione simultanea di 2segnali. LR8514 può misurare2 temperature + 2 umidità.

> > > > > > > > > > >

> > > > > > > > > > >One time

Stop

Out

Sovrascritturasui dati cancellati

Cancellazione dei dati più vecchi

Start recording 500.000 dati/canale

ContinuaLR

85X

X

Page 6: Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione) Display 40 x 25 mm Connettività a Data-Logger tramite Bluetooth® a LR8410/20 Alimentazione

www. .com82

Pulse input cycle

200 μs or higher when the filter is set to OFF(must be 100 μs or higher in H period and L period.)100 ms or higher when the filter is set to ON(must be 50 ms or higher in H period and L period.)

No. of inputchannels

2 channels (common GND)

Measurement modes

Integrating (cumulative/Instant), Revolution, Logic (Records an 1/0 for each recording interval)

Supported input format

Non-voltage “a” contact (always-open contact point), open collector, or voltage input (DC 0 V to 50 V)

Recordingintervals

0.1 to 30 sec, 1 to 60 min, 16 selections

Recording modes Instantaneous value

Dimensions, Weight

85W×61H×31D mm (3.35W×2.40H×1.22D in), 95 g (Not including the battery)

Specifiche tecniche (Accuracy guaranteed for 1 year) Pulse input

LR8512 conta-impulsi & contagiriTotalizza impulsi, misura la velocità di rotazione e controlla stati logici ON/OFF

Registra e gestisci la portata dei fluidiquali acqua, gas, carburanti.Puoi misurare il segnale di uscitadel flussometro (conta-impulsi) evisualizzare le fluttuazioni giornaliere.

Measurement objects

Range Max. Resolution Measurement Range

Totalization 1000M pulse f.s. 1 pulse 0 to 1000 M pulseNo. of revolutions

5000/n [r/s] f.s. 1/n [r/s] 0 to 5000/n [r/s]

* n is the number of pulses, 1 to 1000, per revolution.

MAGNETICSTRAP Z5004

LR8515 tensione + temperaturaMisura tensione e temperatura da termocoppie K e T in simultanea su 2 canali

Manutenzione di impianti fotovoltaici, test su dispositivi elettronici, automobili e mezzi di trasporto.

Con un unico dispositivo puoi misurarela tensione d’uscita in mV da sensori diirraggiamento o di flusso di calore, cosìcome la tensione di batteria.Puoi visualizzare anche la correlazionetra tensione e temperatura.

Specifiche tecniche

No. of input channels

2 ch (isolated; select voltage of thermocouple for each channel)

Measurement items Voltage/Thermocouple (K, T)

Input terminals M3 screw type terminal block (2 terminals per channel)

Maximum input voltage

DC±50 V

Max. inter-channelvoltage

DC 70 V

Recording intervals

0.1 to 30 sec, 1 to 60 min, 16 selections

Recording modes Instantaneous value

Dimensions, Weight

85W×75H×38D mm (3.35W×2.95H×1.50D in), 126 g (Not including the battery)

Specifiche tecniche

Measurement objects

Range Max. Resolution

Measurable Range Measurement Accuracy

Voltage

50 mV f.s. 0.01 mV -50 mV to 50 mV ±0.05 mV500 mV f.s. 0.1 mV -500 mV to 500 mV ±0.5 mV

5 V f.s. 1 mV -5 V to 5 V ±5 mV50 V f.s. 10 mV -50 V to 50 V ±50 mV

Thermo-couples

K 1000 ˚C f.s. 0.1 ˚C-200 ˚C to -100 ˚C ±1.5 ˚C-100 ˚C to 999.9 ˚C ±0.8 ˚C

T 1000 ˚C f.s. 0.1 ˚C-200 ˚C to -100 ˚C ±1.5 ˚C-100 ˚C to 0 ˚C ±0.8 ˚C

0 ˚C to 400 ˚C ±0.6 ˚C

Reference junction compensation: Switchable between internal and externalReference junction compensation accuracy: ±0.5°C (When using internal compensation, add to thermocouple measurement accuracy.)Temperature characteristics: Add (measurement accuracy × 0.1)/°C to measurement accuracy.

Options

CONNECTION CABLE L10101.5 m (4.92 ft)Bundled and also available for additional purchase

AC ADAPTER Z2003100 to 240 VAC, 50/60Hz

MAGNETICSTRAP Z5004

For long-term recording

Options

Accessories: CD-R (Instruction Manual, Logger Utility, Wireless Logger Collector) × 1, Measurement Guide × 1, Caution for Using Radio Waves × 1, AA alkaline batteries (LR6) × 2

Models and accessories

Accessories: CD-R (Instruction Manual, Logger Utility, Wireless Logger Collector) × 1, Measurement Guide ×1, Caution for Using Radio Waves × 1, AA alkaline batteries (LR6) ×2 Connection Cable L1010 × 2

WIRELESS PULSE LOGGER LR8512

*AC Adapter is not included.Models and accessories

WIRELESS PULSE LOGGER LR8515

* Thermocouples and AC Adapter are not included.

AC ADAPTER Z2003100 to 240 VAC, 50/60Hz

Per applicazioni quali:

For long-term recording

Sistemi di condizionamento (flusso dell’aria), automotive (velocità e rotazioni), processi industriali (rotazioni, sensori)

Per applicazioni quali:

Page 7: Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione) Display 40 x 25 mm Connettività a Data-Logger tramite Bluetooth® a LR8410/20 Alimentazione

www. .com 83

LR8513 corrente: CA, CC, dispersioniMisura di corrente in simultanea su 2 canali, con ampia scelta di sensori CA/CC

Assorbimenti e dispersioni di linee elettriche, impianti di sollevamento, ascensori, linee ferroviarie, veicoli elettrici/ibridi.

Per applicazioni quali:

WIRELESS CLAMP LOGGER LR8513

* Current sensor and AC Adapter are not included.

No. of inputchannels 2 channels (common GND)

Measurement items

AC load current, DC load currentAC leak current (using current sensor)

Effective value calculation

Software calculates the true RMS value

Measurement ranges

AC500.0 mA to 2000 A (with current sensor)DC10.00 A to 2000 A (with current sensor)*Current and leak current that occur intermittently cannot be measured.

Measurementaccuracy

±0.5% rdg.±5 dgt. (DC, AC 50/60 Hz)* Add the sensor’s accuracy when the current sensor is connected

Recordingintervals 0.5 to 30 sec, 1 to 60 min, 14 selections

Recording modes Instantaneous value, average value

Dimensions, Weight

85W×75H×38D mm (3. 35W×2.95H×1.50D in) mm, 130 g (Not including the battery)

Specifiche tecniche

Sensor used Range Max. Resolution Measurable Range

9675 500.0 mA 0.1 mA AC 1.0 mA to 500.0 mA5.000 A 0.001 A AC 0.010 A to5.000 A

9657-10 500.0 mA 0.1 mA AC 1.0 mA to 500.0 mA5.000 A 0.001 A AC 0.010 A to 5.000 A

9695-02 5.000 A 0.001 A AC 0.010 A to 5.000 A50.00 A 0.01 A AC 0.10 A to 50.00 A

CT6500 50.00 A 0.01 A AC 0.10 A to 50.00 A500.0 A 0.1 A AC 1.0 A to 500.0 A

9669 1000 A 1A AC 10 A to 1000 A

CT9691-9010.00 A 0.01 A

AC 0.10 ADC± (0.10 A

to 10.00 A to 10.00 A)

100.0 A 0.1 AAC 1.0 ADC± (1.0 A

to 100.0 A to 100.0 A)

CT9692-9020.00 A 0.01 A

AC 0.10 ADC± (0.10 A

to 20.00 A to 20.00 A)

200.0 A 0.1 AAC 1.0 ADC± (1.0 A

to 200.0 A to 200.0 A)

CT9693-90200.0 A 0.1 A

AC 1.0 ADC± (1.0 A

to 200.0 A to 200.0 A)

2000 A 1 AAC 10 ADC± (10 A

to 2000 A to 2000 A)

Current sensor specifications

LR8513 misura il valore istantaneo e il valore medio con un tempo di registrazione a partire da 0.5 secondi.Può quindi rilevare ogni minima variazione di corrente RMS.

Impostando un valore fisso di tensione e fattore di potenza, LR8513 calcola la Potenza Attiva (kWh) sulla base delle variazioni di corrente.

CLAMP ON AC/DC SENSORCT9691-90, CT9692-90, CT9693-90CT9691-90: φ35 mm, AC/DC 100A,

cord length 2 m (6.56 ft)CT9692-90: φ33 mm, AC/DC 200A,

cord length 2 m (6.56 ft)CT9693-90: φ55 mm, AC/DC 2000A,

cord length 2 m (6.56 ft)

AC/DC load current

CONNECTION CABLE 9219For connecting the 9695-02, cord length 3 m (9.84 ft)

CLAMP ONSENSOR CT6500φ46 mm, AC 500A, cord length 3 m (9.84 ft)

AC load currentCLAMP ONSENSOR 9669φ55 mm, AC 1000A, cord length 3 m(9.84 ft)

AC load current

For long-term recording

AC ADAPTER Z2003100 to 240 VAC, 50/60Hz

MAGNETICSTRAP Z5004

Control andcommunications

Bluetooth® 2.1+EDR(Communications range: 30 m, line of sight, security: SSP)

Storage capacity 500,000 data items for each channel

Operating temperature and humidity

Temperature: -20 to 60 ˚C (-4 to 140 °F), Humidity: 80%rh or less (non-condensing)(Depends on battery and current sensor specifications when they are in use)

Storage temperature and humidity

-20˚C to 60˚C, 80%rh or less (non-condensing)(With batteries removed)

Functions

Alarm, Scaling, Recording operation hold function, Erroneous operation prevention, Comment recording function, Power saving function, Authentication function

Display itemsMeasurement value, date, time, number of recorded data, maximum value, minimum value, and average value

Applicable standards

Safety EN61010

EMC EN61326 classA, EN61000-3-2, EN61000-3-3

Wireless certification

Japan: Incorporates a wireless module that has been certified as compliant with applicable technical standards.US: Part 15.247 (Contains FCC ID: QOQWT11IA)Canada:RSS-210 (Contains IC: 5123A-BGTWT11IA)EU: EN 300 328, EN 301 489-1, EN 301 489-17

Vibration endurance JIS D 1601:1995 5.3(1), Category 1: Vehicle, Condition: Category A equiv.

Power source

AC adapter AC ADAPTER Z2003 (sold separately, DC 12 V)

Battery AA alkaline batteries (LR6) × 2

External power

DC 5 V to 13.5 V* can also be supplied from USB bus power, with a conversion cable

Shared specifications LR8512, LR8513, LR8514, LR8515

Accessories: CD-R (Instruction Manual, Logger Utility, Wireless Logger Collector) × 1, Measurement Guide ×1, Caution for Using Radio Waves × 1, AA alkaline batteries (LR6) ×2

Options

Models and accessories

CLAMP ONSENSOR9695-02φ15 mm, AC 50A

Requires the Con-nection Cable 9219

AC load current Insulated conductor

CLAMP ON LEAKSENSOR 9657-10φ40 mm, AC 10A, cord length 3 m (9.84 ft)

Insulated conductor

CLAMP ON LEAKSENSOR 9675φ30 mm, AC 10A, cord length 3 m (9.84 ft)

AC leak current Insulated conductor

LR85

XX

Page 8: Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione) Display 40 x 25 mm Connettività a Data-Logger tramite Bluetooth® a LR8410/20 Alimentazione

www. .com84

LR8514 temperatura + umidità, su 4 canali (2+2)Registra in contemporanea i parametri ambientali di 2 distinti locali

Con LR8514 puoi condurre indaginiad alta efficienza di temperaturae umidità ambientale.Inoltre, dal momento che può registrare contemporaneamente su 2 locali distinti (con 2 sensori abbinati), puoi realizzare analisicomparative ad esempio tra internoed esterno di un particolare ambiente.

Controllo di luoghi di stoccaggio, laboratori, magazzini ad ambiente controllato, celle frigorifere.

Per applicazioni quali:

Measurement objects Range Max. Resolution Measurable Range

Temperature 100 ˚C f.s. 0.1 ˚C -40˚C to 80 ˚C

Humidity 100%rh f.s. 0.1 %rh 0 %rh to 100 %rh

No. of inputchannels

2 ch for temperature + 2 ch for humidity(2 sensors can be attached)

Measurement objects Temperature, Humidity

Temperature measurement accuracy (using Z2010/Z2011)

±0.5˚ C (10 ˚C to 60 ˚C), using Z2010/Z2011If outside above temperature range:Add 0.015 ˚C/ ˚C (-40 ˚C to 10 ˚C) or 0.02˚ C/ ˚C (60 ˚C to 80 ˚C)

Humidity measurement accuracy(using Z2010/Z2011)

±3% rh (20 ˚C to 30 ˚C, 20% to 90% rh)If outside above range, see Figure 1.

Recording intervals 0.5 to 30 sec, 1 to 60 min, 14 selections

Recording modes Instantaneous value

Dimensions, Weight 85W×61H×31D mm (3.35W×2.40H×1.22D in), 95 g (Not including the battery)

Specifiche tecniche

* Only the temperature and humidity sensors affect the measurement accuracy and are subject to calibration.The LR8514 logger does not require calibration.

AC ADAPTER Z2003100 to 240 VAC, 50/60Hz

MAGNETICSTRAP Z5004

Accessories: CD-R (Instruction Manual, Logger Utility, Wireless Logger Collector) × 1, Measurement Guide × 1, Caution for Using Radio Waves × 1, AA alkaline batteries (LR6) × 2

WIRELESS HUMIDITY LOGGER LR8514

* Temperature and humudity sensor, AC Adapter are not included.

HUMIDITY SENSOR Z201050 mm (0.16 ft)

HUMIDITY SENSOR Z20111.5 m (4.92 ft)

OptionsModels and accessories

For long-term recording

±8%rh* ±6%rh* ±8%rh*

±12%rh*

±8%rh*

±6%rh

±3%rh

±6%rh

±6%rh*±6%rh*

±5%rh±5%rh

±5%rh

±3.5%rh ±4%rh

±5%rh

±6%rh ±5%rh

±10%rh* ±8%rh ±4%rh ±6%rh

±12%rh* ±8%rh* ±12%rh*

0 10 20 30 40 50 60 70 80

Temperature[°C]

20

40

60

80

100

0

Rela

tive

hum

idit

y[%

rh]

The accuracy of values indicated by the * mark is not guaranteed (reference values).

Humidity measurement accuracy (fig. 1)

Page 9: Data Logger Multicanale - MYW-CMS€¦ · Software di analisi dati Logger Utility (in dotazione) Display 40 x 25 mm Connettività a Data-Logger tramite Bluetooth® a LR8410/20 Alimentazione

www. .com 85

Specifiche tecniche

* Only the temperature and humidity sensors affect the measurement accuracy and are subject to calibration.The LR8520 logger does not require calibration.

Sebbene l’indice fungino sia un fattore altamente affidabile basato su studi e ricerche accademiche, ciò non esclude che la crescita di funghi e muffe avvenga anche in luoghi con un basso indice fungino.

LR8520 indice fungino in ambienti criticiAnalizza la contaminazione fungina e il fattore di crescita delle muffe

Salvaguardia di alimenti, cereali e documenti da conservare per lunghi periodi nonché per aree espositive, gallerie d’arte, musei.

Per applicazioni quali:

Le spore fungine si diffondono per via aerea e si possono posare su qualsiasi superficie. La loro evoluzione può accelerare in funzione di particolari condizioni ambientali, quali il caldo, l’umido, lo scarso ricambio d’aria, raggruppandosi e trasformandosi rapidamente in muffa. C’è quindi una correlazione diretta tra la crescita fungina ed i valori di temperatura e umidità. Con LR8520 è possibile avere una previsione di sviluppo del fungo, così da attuare le giuste azioni correttive ed evitare la creazione della muffa, che spesso è dannosa ed irreversibile.

MAGNETICSTRAP Z5004

HUMIDITY SENSOR Z201050 mm (0.16 ft)

HUMIDITY SENSOR Z20111.5 m (4.92 ft)

AC ADAPTER Z2003100 to 240 VAC, 50/60Hz

For long-term recordingHUMIDITY SENSOR Z2010 or HUMIDITY SENSOR Z2011 is required to measure humidity.

Supported devices

• Android™ smartphone or Android™ tablet (Download app from Google Play)• Windows PC or Windows tablet (Use bundled software)• WIRELESS LOGGING STATION LR8410-20 *The settings can only be configured from supported devices.

Control andcommunications

Bluetooth® 2.1+EDR or later Profile: SPPCommunications Range: 30 m, line-of-sight(Depends on the performance of the communication supported device)

Display contentsTemperature, humidity, fungal index (0 to 200), growth prediction (5 levels), date, time, number of recorded data, maximum value, minimum value, and average value

Input1 temperature channel + 1 humidity channel(HUMIDITY SENSOR Z2010 or HUMIDITY SENSOR Z2011 is required (sold separately))

Measurable range [Temperature] -40˚C to 80˚C, Range 100˚C f.s., Max. resolution 0.1˚C[Humidity] 0% rh to 100% rh, Range 100% rh f.s., Max. resolution 0.1% rh

Temperature measurement accuracy (using Z2010/Z2011)

±0.5˚ C (10 ˚C to 60 ˚C), using Z2010/Z2011If outside above temperature range:Add 0.015 ˚C/ ˚C (-40 ˚C to 10 ˚C) or 0.02˚ C/ ˚C (60 ˚C to 80 ˚C)

Humidity measurement accuracy (using Z2010/Z2011)

±3% rh (20 ˚C to 30 ˚C, 20% to 90% rh)If outside above range, see Figure 1

Output Outputs alarm signalsRecording intervals 0.5 sec to 30 sec, 1 min to 60 min, 14 selectionsStorage capacity 500,000 data itemsRecording modes Instantaneous value

Continuous record-ing

Set to ON or OFFON: When the storage limit is reached, the oldest data is deleted and the unit continues to record (endless recording)OFF: Recording stops when the storage limit is reached (one-time recording)

Operating temperature and humidity

-20˚C to 60˚C, 80%rh or less (non-condensing)(When operating on battery power, these values will vary according to the battery specifications.)

Storage temperature and humidity

-20˚C to 60˚C, 80%rh or less (non-condensing)(With batteries removed)

Applicable standards

Safety EN61010EMC EN61326 classA, EN61000-3-2, EN61000-3-3

Wireless certification

Japan: Incorporates a wireless module that has been certified as compliant with applicable technical standards.US: Part 15.247 (Contains FCC ID: QOQWT11IA)Canada:RSS-210 (Contains IC: 5123A-BGTWT11IA)EU: EN 300 328, EN 301 489-1, EN 301 489-17

Vibration endurance JIS D 1601:1995 5.3(1), Category 1: Vehicle, Condition: Category A equiv.

Power source

AC adapter AC ADAPTER Z2003 (sold separately, DC 12 V)Battery AA alkaline batteries (LR6) × 2 External power

DC 5 V to 13.5 V* can also be supplied from USB bus power via a conversion cable

Continuous operatingtime (Battery)

3.5 months (Recording interval of 1 min, Bluetooth® OFF)20 days (Recording interval of 1 sec, Bluetooth® ON)

Dimensions and mass

85 mm (3.35 in) W × 61 mm (2.40 in) H × 31 mm (1.22 in) D (Excluding protrusions), 95 g (3.3 oz) (Not including the battery)

±8%rh ±6%rh ±8%rh

±12%rh

±8%rh±6%rh

±3%rh

±6%rh

±6%rh±6%rh

±5%rh±5%rh

±5%rh

±3.5%rh ±4%rh

±5%rh

±6%rh ±5%rh

±10%rh ±8%rh ±4%rh ±6%rh

±12%rh ±8%rh ±12%rh

0 10 20 30 40 50 60 70 80

Temperature[°C]

20

40

60

80

100

0

Rela

tive

hum

idit

y[%

rh]

Humidity measurement accuracy (fig. 1)The accuracy of values shown in is not guaranteed.

(These values are provided only for reference.)

Alarm

ScalingRecording operation hold functionErroneous operationpreventionComment recording functionPower saving functionAuthentication functionFree Run

: ALARM is displayed when values fall outside defined ranges Open drain output (Max. rating: 30 V DC, 200 mA): Measured values are scaled when displayed: If the power source is interrupted while recording, recording will start again automatically when the power is restored.: Confirmation messages are displayed when recording is started/ stopped and when turning the power off: Titles and comments for each channel can be recorded: Power can be saved by turning Bluetooth® on and off as necessary: A password can be set for user authentication: ON/ OFF selection ON: The measurement value is indicated every 1 second while measure-ment is stopped. (the data is not saved in the memory)

The measurement value is saved in the memory every recording interval and indicated every 1 second regardless of recording interval setting while measuring. (when the setting of recording interval is less than 1 second, the measurement value is indicated every recording interval)

OFF: The measurement value which is at the time of a measurement stop is indicated while measurement is stopped.

The measurement value is saved in the memory and indicated every recording interval while measuring.

Functions

LR85

XX