Algoritmi per la delineazione del Biological target volume...
Transcript of Algoritmi per la delineazione del Biological target volume...
diapositiva 1
Marco Brambilla - Roberta Matheoud
Medical Physics Department Azienda Ospedaliero Universitaria “Maggiore della Carità” Novara
Algoritmi per la delineazione del Biological target volume in immagini FDG PET/TC
Incontri a Fisica – Torino 2011
diapositiva 2
METHODS FOR TARGET VOLUME DELINEATION IN PET
1. Visual contouring of PET scan and definition of contours as judged by the experienced physician
2. Absolute Thresholds: Standardized Uptake Value (SUV) >2.5
3. Thresholding by fixed percentage of the maximum uptake or
mean SUV
4. Algorithms depending on lesion size 5. Algorithms depending on contrast 6. Complex algorithms
diapositiva 3
1. Visual Contouring
Author Reference District Patient n°
Faria SL et al Int J Radiat Oncol Biol Phys 2008;79:1035-8 Lung 32
Schwartz D et al. Head Neck 2005;27:478-487 Head Neck 63
Heron DE et al. Int J Radiat Oncol Biol Phys 2004;60:1419-1424 Head Neck 21
Nestle U et al Int J Radiat Oncol Biol Phys 1999;44:593-597 Lung 34
Kiffer JD et al Lung Cancer 1998;19:167-177 Lung 15
This method is the first one applied and still widely used Easily applicable from a technical point of view
METHODS FOR TARGET VOLUME DELINEATION IN PET
But …
diapositiva 4
Riegel et al reported significant differences in GTV delineation between multiple observers (A,B,C,D) contouring on PET/CT fusion. Thus, confirming the need for a delineation protocol.
Riegel et al. Int J Radiat Oncol Biol Phys 2006;65:726-732
A B C D
A - 24% 12% 90%
B 15% - 12% 79%
C 45% 56% - 137%
D 5% 8% 1% -
METHODS FOR TARGET VOLUME DELINEATION IN PET
… is affected by a significant interobserver variability
diapositiva 5
2. Absolute SUV (GTV= SUV>2.5)
Author Reference District Patient N°
Paulino AC et Int J Radiat Oncol Biol Phys 2004; 59:4-5 -----
Nestle U et al J Nucl Med 2005;46:1342-1348 Lung 25
Nestle U et al. Eur J Nucl Med Mol Imag 2007;34:453-462 Lymphnodes 15
Well applicable to most PET and to some RT planning systems
Due to biological and physical factors there are no “normal” values for SUVs to be similarly used in every case.
It has been shown that this method often fails, e.g. when the physiological background activity lies above the fixed threshold (Nestle et al 2007)
METHODS FOR TARGET VOLUME DELINEATION IN PET
diapositiva 6
2. Absolute SUV
18F-FDG PET: SUVmax = 30 Isocontours: narrow = GTV40 wide = GTV2.5
Nestle et al. (2005) concluded that, as for NSCLC: 1. GTV40 does not appear to be suitable for target volume delineation. 2. GTV2.5 provided volumes higher than GTVCT and similar to those
provided by visual interpretation
METHODS FOR TARGET VOLUME DELINEATION IN PET
Nestle et al. (2005): Corresponding planning CT: red = GTV40, green = GTVbg, yellow = GTVCT.
diapositiva 7
Many factors affect SUV measurements and therefore tumor contours:
– The metabolic activity of a tumor, the glucose level in the blood,
the time elapsing from injection to acquisition, the reconstruction
algorithm.
– The lesion dimensions (partial volume effect)
– Heterogeneity within a tumor
– Tumor motion
The use of a fixed SUV threshold sorrounding the lesion is not suitable for GTV contouring
METHODS FOR TARGET VOLUME DELINEATION IN PET
2. Absolute SUV
diapositiva 8
3. Fixed percentage of mean SUV
Variable threshold for SUVmean: Threshold SUV= 0.307 x SUVmean + 0.588
Author Reference District Patient N° Black QC et al. Bayne M et al
Int J Radiat Oncol Biol Phys 2004; 60:1272-1282 Int J Radiat Oncol Biol Phys 2004; 60:1272-1282
Lung Lung
15 26
SUVmean depends on ROI dimensions At least iterative process
Circular reasoning
Bayne et al reported discrepant GTVs using this algorithm
to evaluate patient data
ROI must be defined to calculate SUVmean
METHODS FOR TARGET VOLUME DELINEATION IN PET
diapositiva 9
3. Fixed %Threshold
A few early investigations estimated that a threshold of ~ 40% of maximum uptake approximated tumor volume
Author Reference District Patient N° Erdi YE et al. Radiother Oncol 2002; 62:51-60 Lung 11
Erdi YE et al. Cancer 1997;80:2505-2509 Lung 10
Many clinical studies were undertaken adopting a fixed threshold of maximum uptake comprised between 40% and 50%:
Author Reference District Patient N° Method
Bassi MC et al Int J Radiat Oncol Biol Phys 2007 Rectum 25 FT 42%
Koshy M et al. Head neck 2005;27:494-502 HN 36 FT 50%
Scarfone et al. J Nucl Med. 2004 Apr;45:543-52 HN 6 FT 50%
Bradley et al Int J Radiat Oncol Biol Phys. 2004;59:78-86 Lung 26 FT 40%
Ciernik IF et al Int J Radiat Oncol Biol Phys 2003;57:853-863 HN Lung
Pelvis 39 FT 50%
METHODS FOR TARGET VOLUME DELINEATION IN PET
diapositiva 10
This method is not likely to be accurate over the full range of clinically relevant volumes and target contrast, mainly because:
1. Thresholds depend on target size 2. Thresholds depend on target contrast 3. Thresholds are “scanner specific” since they likely depend on the
spatial resolution characteristics of the scanner used
GTV delineation of primary tumors, tresholding by 40% of the maximum accumulation intensity should be avoided due to the risk of insufficient tumor coverage (Nestle et al 2009).
METHODS FOR TARGET VOLUME DELINEATION IN PET
3. Fixed %Threshold
diapositiva 11
Erdi YE et al. Cancer 1997;80:2505-2509
METHODS FOR TARGET VOLUME DELINEATION IN PET
Drever L et al. Med Phys 2007;34:1253-1265
4. Algorithms depending on lesion size
Ford EC et al. Med Phys 2006;33:4280-4288
C-PET Plus (Philips)
diapositiva 12
Jentzen W et al. J Nucl Med 2007;48:108-114 Daisne JF et al. Radiother Oncol 2003;69:247-250
5. Algorithms depending on contrast
ECAT HR (Siemens)
Th = a + b/(S/B)
Nestle U et al. J Nucl Med 2005;46:1342-1348
Imean= mean intensity of all pixels surrounding the 70% Imax isocontour within the tumor
ECAT Art (Siemens)
ITh = (0.15xImean) + B
METHODS FOR TARGET VOLUME DELINEATION IN PET
diapositiva 13
6. Complex algorithms: Background Subtracted Images
Thabs=B + Threl(Smax-B)
Threl= 41% for diameter > 12 mm
ECAT Exact HR+ (Siemens)
Davis JB et al. Radiother Oncol 2006;80:43-50
METHODS FOR TARGET VOLUME DELINEATION IN PET
diapositiva 14
6. Complex algorithms: Iterative Image Thresholding
Discovery LS (GE)
Jentzen W et al. J Nucl Med 2007;48:108-114
METHODS FOR TARGET VOLUME DELINEATION IN PET
Th(%) = 7.8%/V(mL) + 61.7%*B/S + 31.6%
diapositiva 15
6. Complex algorithms: Gradient based segmentation
Ecat HR Plus (Siemens)
Geets X et al. E J Nucl Med Mol Imaging 2007;34:1427-1438
a. Image denoising b. Image deblurring c. Gradient based segmentation
METHODS FOR TARGET VOLUME DELINEATION IN PET
diapositiva 16
Conclusions
• Visual contouring is affected by a significant interobserver variation and should be avoided.
• The use of a fixed SUV threshold surrounding the lesion is not suitable for GTV contouring. The method often fails when the physiological background activity lies above the fixed threshold.
• The use of a fixed threshold of maximum uptake (e.g 40%) should be avoided due to the risk of insufficient tumor coverage.
• Methods for segmentation of non uniform tracer concentration not based on thresholding have been recently proposed. While referring to these promising methods it should be pointed out that until they are further developed and validated, contrast-oriented segmentation methods are and will be used in most clinics and therefore need to be accurately characterized
METHODS FOR TARGET VOLUME DELINEATION IN PET
diapositiva 17
Imaging parameters not explored in contrast oriented methods
• Effects of Image Acquisition Parameters – Emission scan duration and background activity concentration are both
related to noise in the reconstructed image. Their effects on TS was not ascertained.
– If different levels of noise play a role in threshold determination, the calibration curves will be specific for the acquisition protocol
• Effects of attenuation and scatter – If attenuation and scatter play a role in threshold determination, site
specific thresholds should be devised. – Threshold values are usually determined using simple phantoms
(cylinder/rectangular box with spherical/rod sources). The results obtained in these simplified conditions may not be reliable in more realistic conditions of attenuation and scattering.
• Effects of Image Reconstruction parameters – More smoothing require higher fractional TS – The role of the convergence of iterative algorithms (i.e iteration number)
was not ascertained.
diapositiva 18
Imaging parameters not explored in contrast oriented methods
• Activity distribution – The published thresholding methods provide reliable volume estimates only
if the imaged activity distribution is homogeneous. • Lesion movement
– Errors associated with lesion masses moving during data acquisition, (e.g. imaging of lung lesions), are unpredictable and not yet characterized.
• Spherical vs irregular shape – Most of the published thresholding methods assumed spherical lesions and
the clinical PET volumes calculations are performed using an ellipsoid model. These suppositions are approximation of the irregularly shaped tumors.
diapositiva 19
A multivariable approach was adopted to study the dependence of the percentage threshold (TH (%)) used to define the boundaries of 18F-FDG positive tissue on emission scan duration (ESD) and activity at the start of acquisition (Aacq) for different target sizes (Sphere ID) and target-to-background (T/B) ratios.
AIM OF THE STUDY
Med Phys 2008;34:1253-1265
diapositiva 20
MATERIALS AND METHODS
• 6 fillable spheres ID: 10, 13, 17, 22, 28 and 37 mm • 4 micro hollow spheres ID: 4.1, 4.7, 6.5, 8.1 mm • 3 cm-thick annular ring of water bags fit over IEC
phantom to match T+S rates estimated for patients
• Polyethylene cylinder (ρ=0.96 g/cm3, Ø = 203 mm, length = 700 mm)
• fillable 80 cm-long plastic tube (ØID = 3.2 mm) parallel to the central axis of the cylinder at a 45 mm radial distance.
SCANNER
TEST PHANTOM SET IEC body phantom + water bags Scatter phantom
The phantom sets used well reproduce clinical count rates.
Scanner: Biograph 16 Hi-Rez PET/CT (Siemens)
diapositiva 21
SOURCE PREPARATION
Torso cavity: 12 kBq/mL Spheres (ID=28 and 37 mm): non radioactive water Other spheres: activity concentrations that provided T/B ratios of 23, 10 and 4 Scatter phantom to simulate activity outside FOV (12 kBq/ml in plastic tube) ESD: 1, 2, 3 and 4 min/bed. Repeated acquisitions to have PET images with varying background activity concentrations in the IEC phantom of about 12, 9, 6, 5 and 3 kBq/mL.
IMAGE ACQUISITION
The results are reported for reconstructions using the attenuation weighted-FORE-OSEM iterative reconstruction, with 2 iterations - 8 subsets and 4 mm FWHM Gaussian filter. The data were reconstructed over a 256x256 matrix with 2.625 mm pixel size and 2 mm slice thickness.
IMAGE RECONSTRUCTION
THRESHOLD DETERMINATION The thresholds were determined as a percentage of the maximal activity concentration in the spheres (TH (%)). Agreement was deemed to occur when volumes (measured versus physical) differed by less than 3%.
diapositiva 22
STATISTICAL ANALYSIS T/B, Aacq, ID, ESD TH (%)
only on visible spheres
multiple linear regression
Data were fitted separately for Sphere ID ≤ 10 mm and Sphere ID > 10 mm
Model Fitting
EESDBSphereIDBABratioBT
BBTH acq +×+×+×+−×+= 43210 )/
11((%)
The F statistic was used as a criterion for selecting a model, being a higher F value associated with a smaller residual sum of square.
diapositiva 23
SphereID.)TB
((%)TH ×−−×−= 6911172309
Neither ESD, nor Aacq resulted as statistically significant predictors of TH (%)
T/B ratio
T/B ratio
Sphere ID (mm)
Sphere ID (mm)
Sphere ID > 10 mm
SphereID.)TB
((%)TH ×−−×−= 98011101151
Sphere ID (β3= - 0.94) T/B ratio (β1= -0.55)
T/B ratio (β1= -0.88) Sphere ID (β3= - 0.32)
Multiple R2 = 0.86
Multiple R2 = 0.86
Sphere ID ≤ 10 mm
diapositiva 24
• The model fittings of TH (%) are good, accounting for more than 86% of TH (%) variance.
• Neither ESD, nor Aacq resulted as statistically significant predictors of TH (%)
• The relevant predictors of TH (%) are Sphere ID and T/B ratio • The most relevant contributors to TH (%) variance are Sphere ID for
Sphere ID ≤ 10 mm and T/B ratio for Sphere ID >10 mm • Approaches which focus on determining a single value for the threshold
contour level that returns the actual volume are not likely to depict the actual spectrum of clinically relevant volumes and T/B ratios. Both target size and T/B ratios play a major role in explaining the variance of TH (%), throughout the whole range of target sizes and T/B ratios examined. Thus, algorithms aimed at automatic threshold segmentation should incorporate both variables with a relative weight which critically depends on target size.
DISCUSSION AND CONCLUSIONS
diapositiva 25
To evaluate the role of different amount of attenuation and scatter on FDG-PET image segmentation using an adaptive threshold method based on the target-to-background (TB) ratio and target dimensions.
AIMS OF THE STUDY
To first show that also with simplified phantom geometries as those selected in this study, the accuracy of scatter and attenuation correction decreases with increasing amount of attenuation and scatter in the phantom being imaged
To evaluate the possibility of defining a unified contrast-based method for target volume delineation in different anatomical districts
diapositiva 26
TEST PHANTOM SET
IEC + 3 cm water bags Hoffman 3D
Scatter
IEC + 9 cm water bags
MATERIALS AND METHODS
diapositiva 27
STATISTICAL ANALYSIS
TB, sphere A
TS (%) Multiple regression
Covariates
Phantom
Factor
diapositiva 28
STATISTICAL ANALYSIS
Starting Model
ETB
BmmAsphereBB +−×+×+= )11()(Y 22
10
EZXBZXBZXBZXBZBZBXBXBBY +++++++++= 228217126115241322110
( ) ( ) EXBBXBBBBY ++++++= 26215130
( ) ( ) EXBBXBBBBY ++++++= 28217140
Suppose now that we wish to compare the separate multiple regressions of TS on A and (1-1/TB) for the three phantoms. For each case in each phantom group, we observe values of the variable Y=TS, X1=A and X2=(1-1/TB); further, we shall suppose that there are ni cases in the ith Phantom group, i=1,2,3. We begin by defining two dummy variables Z1 and Z2 as follows: Z1=1 if Phantom 2 case; 0 otherwise; Z2=1 if Phantom 3 case; 0 otherwise. The complete model to be used is then given as follows:
For each particular Phantom, model (3) specializes as follow: Phantom 1 (Z1=Z2=0) Phantom 2 (Z1=1;Z2=0)
EXBXBBY +++= 22110
Phantom 3 (Z1=0;Z2=1)
diapositiva 29
STATISTICAL ANALYSIS
The following hypotheses concerning the parameters in model are of interest: Are all the three regression equations coincident (i.e, test H0: B3=B4=B5=B6=B7=B8=0)? When H0 is true, all three phantom models reduce to the form:
EXBXBBY +++= 22110
The test statistic is the multiple partial F:
[ ])mod(
/)mod()mod(elfullresidualMS
elfullSSresidualelreducedSSresidualF
ν−=
where: residual SS = residual sum of squares; MS = mean square; n = number of independent linear parametric functions specified to be 0 under H0. H0 must be tested using Analysis of Variance tables with n =6 and n1+n2+n3-ñ degrees of freedom, being ñ the number of parameters in the full model.
diapositiva 30
RESULTS
Table 2. Physical properties of the phantoms.
Phantom set
Scatter Fraction Peak NECR (k=2) Accuracy of attenuation and
scatter correction (DClung)
Phan 1 27.4 % 136.6* kcps at 70.9 kBq/ml
23.7 %
Phan 2 40.4 % 61.2 kcps at 31.8 kBq/ml
37.6 %
Phan 3 48.7 % 30.8* kcps at 25.2 kBq/ml
40.8 %
* Peak NECR was not reached due to insufficient starting activity
diapositiva 31
Hoffman IEC + 3 cm w. IEC + 9 cm w.Phantom
20
22
24
26
28
30
32
34
36
38
40
42
44
Err
or in
sca
tter a
nd a
ttenu
atio
n co
rrec
tion
(%)
Also with simplified phantom geometries as those selected in this study, the accuracy of scatter and attenuation correction decreases with increasing amount of attenuation and scatter in the phantom being imaged.
100,
,, ×=∆
ibkgd
ilungilung c
cC
RESULTS
Errors in scatter and attenuation correction (%)
Overall, this increase is statistically significant (F= 406.8 p <10-6). Individual significant differences (p=<10-4) were found between Phantom 1 (∆Clung=23.7 ± 1.6 %) and Phantom 2 (∆Clung=37.6 ± 3.4 %) and between Phantom 2 and Phantom 3 (∆Clung=40.8 ± 5.1 %) (p=<10-4).
diapositiva 32
RESULTS
The TS versus cross sectional area relationship was splitted and fitted into different functional forms as already performed in previous studies. Partial volume effects significantly reduce the contrast recovery for structures less than three-times the reconstructed image resolution which in our scanner is about of 4.5 mm. Thus the choice of sphere A=133 mm2 (or, equivalently, a sphere internal diameter of 13 mm) as a separator of the data was dictated by the resolution characteristics of our scanner.
Cross section A ≤ 133 mm2 )11(1.63)(20.07.127(%) 2TB
mmsphereATS −×−×−=
The adjusted coefficient of determination is R2 =0.80. Both sphere A and (1-1/TB) resulted as statistically significant predictors of TS. TS diminish with increasing sphere A and with increasing TB ratios. The functional dependence of TS on (1-1/TB) is higher (β (1-1/TB) = -0.67) than the influence of A (βA=-0.55) for the small volumes corresponding to sphere A≤133 mm2. The test of the hypothesis of coincident regression lines for the three phantoms provided an F 6, 13 = 0.09 (P= 0.996). This F statistic is extremely small (P is very large); so we do not reject H0 and therefore have no statistical basis for believing that the three lines are not coincident.
diapositiva 33
RESULTS
Analysis of Variance table. Test of H0 = coincident regression lines for the three phantoms. A≤133 mm2
Reduced Model
Sum of squares (SS) Degrees of freedom
Mean Square (MS)
F p
Regression 3061.8 2 1530.9 43.7 0,000000
Residuals 665.1 19 35.0
Full Model
Sum of squares (SS) Degrees of freedom
Mean Square (MS)
F p
Regression 3081.9 4 770.5 20.3 0,000003
Residuals 645.0 17 37.9
[ ] 996.009.09.37
6/0.6451.66513,6 ==
−= PF
diapositiva 34
RESULTS
Cross section A > 133 mm2 )11(6.622.95(%)TB
TS −×−=
The adjusted coefficient of determination is R2 =0.73. Only (1-1/TB) resulted as a statistically significant predictor of TS, while sphere A is no longer retained into the model. Again TS diminish with increasing TB ratios. The test of the hypothesis of coincident regression lines for the three phantoms provided an F4, 30 = 1.03 (P= 0.408). This F statistic is small (P is large); so we do not reject H0 and therefore have no statistical basis for believing that the three lines are not coincident.
diapositiva 35
RESULTS
Analysis of variance table. Test of H0 = coincident regression lines for the three phantoms. A> 133 mm2
Reduced Model
Sum of squares (SS)
Degrees of freedom
Mean Square (MS)
F p
Regression 1178.3 1 1178.3 95.5 0,000000
Residuals 1942.6 34 12.3
Total 1647.8
Full Model
Sum of squares (SS)
Degrees of freedom
Mean Square (MS)
F p
Regression 1226.3 3 408.8 35.2 0,000000
Residuals 371.2 32 11.6
Total 1597.6
[ ] 407.003.16.11
4/2.3713.41930,4 ==
−= PF
diapositiva 36
• Different conditions of attenuation and scatter do not play a significant role in explaining the variance of TS the wide range explored of TB ratios, target sizes and reconstruction parameters.
• This, in turn ensures that calibration curves needed to implement contrast-based contouring algorithms of PET volumes can be devised in different conditions of attenuation and scatter.
• This, together with the previously demonstrated independence of TS on acquisition parameters such as emission scan duration or background activity concentration adds to the generalizability of this methods: distinct calibration curves need not to be specifically devised for each anatomical site representing different conditions of attenuation and scatter and, on the other hand, curves generated in one PET center for well defined conditions can be exported to other centers using different working conditions in terms of emission scan duration, administered activities or delay between injection and acquisition, provided that both are equipped with the same scanner and use the same reconstruction parameters.
• This opens the possibility of defining a unified contrast-based method for target delineation in different anatomical districts.
DISCUSSION AND CONCLUSIONS
diapositiva 37
To select the reconstruction scheme which maximizes the goodness of fit of the calibration curve and the robustness of the thresholding algorithm .
AIMS OF THE STUDY
To analyze the behaviour of a contouring algorithm for PET images based on adaptive thresholding depending on lesions size and target-to-background (TB) ratio under different conditions of image reconstruction parameters.
diapositiva 38
TEST PHANTOM SET IEC + 3 cm water bags Scatter
Image reconstruction parameters
16 iterations 4 mm FWHM
32 iterations 4 mm FWHM
64 iterations 4 mm FWHM
16 iterations 6 mm FWHM
32 iterations 6 mm FWHM
64 iterations 6 mm FWHM
16 iterations 8 mm FWHM
32 iterations 8 mm FWHM
64 iterations 8 mm FWHM
Convergence
Smoothing
MATERIALS AND METHODS
diapositiva 39
Images of different targets with internal diameters ranging from 10 to 37 mm and acquired with a source-to-background ratio of 8.1.
Changes in image appearance of different targets as number of iterations is increased from 16 (top row) to 64 (bottom row) and as smoothing kernel is increased from FWHM=4 mm (left column) to FWHM= 8 mm (right column). The data show progressively noisier images but with less smoothing and more spatial features as the number of iterations increases.
The amount of Gaussian post-reconstruction smoothing was chosen so to span a resolution range in the reconstructed images of at least twofold the value measured with filtered back projection.
diapositiva 40
RESULTS
30
40
50
60
70
80
90
100
0 200 400 600 800 1000 1200
A (mm2)
TS (%
)
TB2
TB3
TB4
TB5-7
TB8-14
TB15-22
TB23-33
TB>33
Example of TS required to produce correct target delineation over the full range of the sphere cross sectional areas examined. The dependence upon TB ratio is also represented in this graph. PET reconstructions were accomplished with iterative algorithm of OSEM obtained with 8 subsets and 2 iterations with a Gaussian smoothing filter with a width of 8 mm. Different reconstruction strategies exhibit similar trends.
diapositiva 41
RESULTS Example of TS required to produce correct target delineation over the full range of the sphere cross sectional areas examined. The dependence upon TB ratio is also represented in this graph. PET reconstructions were accomplished with iterative algorithm of OSEM obtained with 8 subsets and 2 iterations with a Gaussian smoothing filter with a width of 8 mm. Different reconstruction strategies exhibit similar trends.
diapositiva 42
RESULTS Example of TS required to produce correct target delineation over the full range of the sphere cross sectional areas examined. The dependence upon TB ratio is also represented in this graph. PET reconstructions were accomplished with iterative algorithm of OSEM obtained with 8 subsets and 2 iterations with a Gaussian smoothing filter with a width of 8 mm. Different reconstruction strategies exhibit similar trends.
diapositiva 43
RESULTS
Cross section A ≤133 mm2
Both sphere A and (1-1/TB) resulted as statistically significant predictors of TS for each reconstruction algorithm examined. TS diminish with increasing sphere A and with increasing TB ratios. With the exception of 2i8s4mm algorithm, the functional dependence of TS on TB is lower than the influence of cross sectional area for the small volumes corresponding to sphere A≤133 mm2.
)11()((%) 22
10 TBBmmsphereABBTS −×+×+=
diapositiva 44
Convergence
Smoothing
16 iterations – 4 mm FWHM R2 = 0.91 16 iterations – 6 mm FWHM R2 = 0.91 16 iterations – 8 mm FWHM R2 = 0.90 32 iterations – 4 mm FWHM R2 = 0.77 32 iterations – 6 mm FWHM R2 = 0.91 32 iterations – 8 mm FWHM R2 = 0.87 64 iterations – 4 mm FWHM R2 = 0.74 64 iterations – 6 mm FWHM R2 = 0.87 64 iterations – 8 mm FWHM R2 = 0.87
Goodness of fit
RESULTS
Only a slight diminishing trend can be observed with increasing EM equivalent iteration number
No clear tendency emerges as for the amount of Gaussian smoothing.
Coefficient of determination
diapositiva 45
ANALYSIS OF COVARIANCE
Thresholds are higher when more smoothig is applyed
FWHM = 4 mm FWHM = 6 mm FWHM = 8 mm
Smoothing
56
58
60
62
64
66
68
70
72
74
TS (%
)
Smoothing
sixteen thirtytwo sixtyfour
Iterations
56
58
60
62
64
66
68
70
72
74
TS (%
)
Convergence
Different conditions of convergence during OSEM reconstruction do not influence threshold determination
diapositiva 46
Regression equation with parameters of reconstruction
• The regression equation that best summarizes the results obtained pooling all the data in a multiple regression model with TS as predicted variable and TB ratio, sphere A and amount of post-reconstruction Gaussian smoothing as predictor variables may be written as:
• The adjusted coefficient of determination is R2 =0.88. The variables inserted into the model were statistically significant predictors of TS, whose variance can be accounted for, in order of decreasing relevance, by target dimensions (βA= -0.64), contrast (β(1-1/TB)= -0.52), and smoothing (βFWHM= 0.29).
)(82.2)11(96.56)(25.09.112(%) 2 mmFWHMTB
mmsphereATS ×+−×−×−=
diapositiva 47
RESULTS
Cross section A > 133 mm2
Only (1-1/TB) resulted as statistically significant predictors of TS, while sphere A is no longer retained into the model as a significant predictor for each reconstruction algorithm examined.
)11((%) 10 TBBBTS −×+=
diapositiva 48
Convergence
Smoothing
16 iterations – 4 mm FWHM R2 = 0.91 16 iterations – 6 mm FWHM R2 = 0.95 16 iterations – 8 mm FWHM R2 = 0.94 32 iterations – 4 mm FWHM R2 = 0.74 32 iterations – 6 mm FWHM R2 = 0.88 32 iterations – 8 mm FWHM R2 = 0.92 32 iterations – 8 mm FWHM R2 = 0.90 64 iterations – 6 mm FWHM R2 = 0.82 64 iterations – 8 mm FWHM R2 = 0.90
Goodness of fit
RESULTS
Only a slight diminishing trend can be observed with increasing EM equivalent iteration number
The increase of Gaussian smoothing increases the goodness of fit of the selected reconstruction algorithm.
Coefficient of determination
diapositiva 49
ANALYSIS OF COVARIANCE
Thresholds are higher when more smoothing is applyed
Smoothing
Convergence
Different conditions of convergence during OSEM reconstruction do not influence threshold determination
FWHM = 4 mm FWHM = 6 mm FWHM = 8 mm
Smoothing
40,0
40,5
41,0
41,5
42,0
42,5
43,0
43,5
44,0
44,5
45,0
45,5
46,0
TS (%
)
sixteen thirtytwo sixtyfour
Iterations
40,0
40,5
41,0
41,5
42,0
42,5
43,0
43,5
44,0
44,5
45,0
45,5
46,0
TS (%
)
diapositiva 50
Regression equation with parameters of reconstruction
• The regression equation that best summarizes the results obtained pooling all the data in a multiple regression model with TS as predicted variable and TB ratio and amount of post-reconstruction Gaussian smoothing as predictor variables may be written as
• The adjusted coefficient of determination is R2 =0.87. The variables inserted into the model were statistically significant predictors of TS, whose variance can be accounted for, in order of decreasing relevance, by contrast (β(1-1/TB)= -0.91), and smoothing (βFWHM= 0.13).
)(60.0)11(27.6040.93(%) mmFWHMTB
TS ×+−×−=
diapositiva 51
Discussion
The results of our study show that the degree of convergence of OSEM algorithms does not influence TS determination for target cross sectional areas lower than 133 mm2, while only a slight tendency toward an increase of TS with increasing iterations can be observed for bigger targets. However, this is always a second-order effect in comparison to smoothing. No significant interaction was detected between the number of iterations and the amount of smoothing. This brings together another relevant consequence: the number of iterations and the amount of smoothing can be optimized independently. The inclusion of smoothing in the adaptive thresholding algorithm adds to the generalization of the method allowing its use in centres equipped with the same scanner but using a different set of reconstruction parameters.
diapositiva 52
Target definition of moving lung tumors in positron emission tomography: Correlation of optimal activity concentration thresholds with object
size, motion extent, and source-to-background ratio
AIMS OF THE STUDY
To develop a functional relationship between optimal activity concentration threshold, tumor volume, motion extent, and SBR using multiple regression techniques by performing an extensive series of phantom scans simulating tumors of varying sizes, SBR, and motion amplitudes.
Riegel AC et al. et al. Med Phys 2010;37:1742-1752
CONCLUSION
lnw = B1x1/3 + B2y1/3 + B3 ln(z )+ B4(x1/3 ln(z)) + B5(y1/3 ln(z)) + B6(xy)1/3 + B7.
where w is the threshold normalized to background, x is the volume in cm3, y is the motion in mm, z is the SBR which is unitless, and Bn are the regression coefficients.
diapositiva 53
Conclusions
Adaptive threshold algorithms do not depend on ESD or background activity concentration.
Image contrast is not the only determinant of threshold since also target dimension play an independent role at least for small targets as can be found when contouring head and neck tumours or lymphnodes.
Iterative methods are necessary since image contrast can be measured but target dimensions can only be guessed.
Different conditions of image noise do not play any role in TS determination, at least in the range explored.
diapositiva 54
Conclusions
Adaptive threshold algorithms do not depend on scatter and attenuation conditions
Distinct calibration curves need not to be specifically devised for each anatomical site representing different conditions of attenuation and scatter and, on the other hand, curves generated in one PET center for well defined conditions can be exported to other centers using different working conditions in terms of emission scan duration, administered activities or delay between injection and acquisition, provided that both are equipped with the same scanner and use the same reconstruction parameters.
The degree of convergence of OSEM algorithms does not influence TS determination
The inclusion of smoothing in the adaptive thresholding algorithm adds to the generalization of the method allowing its use in centres equipped with the same scanner but using a different set of reconstruction parameters.
diapositiva 55
Validation of the algorithm
• N= 24 patients with 33 lymphnodes examined • We excluded from the analysis mediastinal lymphnodes.
Breathing excursion do not play a role. • Lymphnodes are spherical and their dimensions are known from
co-registered CT. • Unlike the large and often inhomogeneous primary tumours,
limph nodes are mostly small and FDG uptake appears rather homogeneous.
diapositiva 56
Patient CA Lymphnode:
CT Volume: 0.9 mL
PET Volume: 0.8 mL
CASE 1
diapositiva 57
Slice Volume TC (mL)
Area TC (mm2)
TB TS % Volume PET (mL)
139 0,23 46 2,6 88 0,21
140 0,36 72 2,9 80 0,36
141 0,26 52 2,1 90 0,23
diapositiva 58
Patient DG Lymphnode: Inguinal sx
CT Volume: 3.2 mL
PET Volume: 2.9 mL
CASE 2
diapositiva 59
Slice Volume CT (mL)
Area CT (mm2)
TB TS % Volume PET (mL)
36 0,25 50 1,8 94 0,15
37 0,28 56 4,2 78 0,25
38 0,6 120 9,9 56 0,53
39 0,88 176 12,8 42 0,76
40 0,67 134 6,2 48 0,76
41 0,48 96 3,0 73 0,46
3,16 2,91
CASE 2
diapositiva 60
CASE 3 Patient DR Lymphnode: Ascilla DX
CT Volume: 7.6 mL
PET Volume: 6.3 mL
diapositiva 61
CASE 3
Slice Volume CT (mL)
Area CT (mm2
)
TB Soglia Volume PET (ml)
108 0,75 150 2,8 61 1,14
107 0,77 154 3,1 59 0,59
106 1,32 264 3,8 55 0,74
105 1,33 266 3,9 54 0,92
104 1,44 288 4,1 53 0,95
103 1,27 254 5,1 50 1,07
102 0,74 148 3,7 55 0,9
7,62 6,31
diapositiva 62
Plot of differences in Volume for all GTVs vs lesion size
-1,6
-1,2
-0,8
-0,4
0,0
0,4
0,8
1,2
1,6
0 2 4 6 8 10 12 14
volume of lesions derived from CT (mL)
diffe
renc
e fro
m C
T in
vol
ume
(mL)
diapositiva 63
Algoritmi dipendenti dal contrasto
Conclusioni II
Durata acquisizione emissiva Concentrazione attività di background
Attenuazione e scatter
Parametri di ricostruzione: N°iterazioni, smooth
Dimensione della lesione
CNR lesione
soglia = f CNR dimensione smooth
Questo risultato apre la possibilità di definire un unico algoritmo
per la delineazione dei volumi per diversi distretti anatomici, non dipendendo dalle condizioni di attenuazione e scatter
diapositiva 64
Studio multicentrico soglie
Prospettive
AO Niguarda Ca' Granda, MI Istituto Europeo di Oncologia, MI
Fondazione S.Raffaele Monte Tabor, MI A.O. San Gerardo, MB
AOU Maggiore della Carità, NO A.O. Arcispedale S.Maria Nuova, RE
A.O. San Camillo Forlanini, RM AOU Integrata di VR
F.P.O. IRCC Candiolo, TO
‘Multi-soglie’
Questo risultato vale su tutti i tomografi PET ?
Discovery STE, GE
Discovery DST, GE
Discovery 600, GE
Biograph 6 True V, Siemens
Gemini TF TOF 16, Philips
Biograph 16 Hirez-pico3D, Siemens
diapositiva 65
+ microsfere + sferona (~100ml)
Saranno realizzati 9 TB, per ogni TB il fantoccio verrà acquisito
a 4 tempi/bed diversi, ogni acquisizione verrà ricostruita in 6
modi diversi (variando iterazioni e smooth):
9 TB x 4acq/TB x 6 ric/TB
216 set di dati per ogni centro
soglia = f CNR dimensione smooth