    PatientPatientpositioning andpositioning and

    centering in CTcentering in CTEtienneEtienne GrechGrech

    CT Unit, Mater Dei Hospital.CT Unit, Mater Dei Hospital.

  • 7/29/2019 centraggio in ct

    2/25

    Aim of presentationAim of presentation To differentiate between the term

    positioning and centering.

    To determine the effects on patientdose and image quality of impropercentering and positioning.

  • 7/29/2019 centraggio in ct

    3/25

  • 7/29/2019 centraggio in ct

    4/25

    Radiation dose & reductionRadiation dose & reduction

    techniquestechniques Radiation doses in CT range between 1-

    24mSv for adults and 2 to 6.5mSv. Heightened concerns and awareness ofradiation induced cancer from CTprompted to the development of

    strategies and techniques for dosereduction. In order to obtain appropriate benefits

    from these techniques, it is important toappropriately centre/position patients tothe isocenter of the gantry.

  • 7/29/2019 centraggio in ct

    5/25

    Positioning and centeringPositioning and centering Positioning: the proper placement of

    the patient on the CT table in the z-axis or along the length of patient forscanning an anatomical region.

    Centering: the appropriateplacement of the patient with respect

    to the gantry isocenter in the x-yplane.

  • 7/29/2019 centraggio in ct

    6/25

    Patient positioningPatient positioning Improper positioning: may result in the

    acquisition of a reference radiograph beyond theROI and/or require a repeat of the latter.

  • 7/29/2019 centraggio in ct

    7/25

    PositioningPositioning cont.cont.

    Proper positioning in head

    examinations is important so as toadequately ensure the reproductionof any asymmetry due to anypathological process.

    Avoiding the eyes by depressing the

    patients chin may reduce lens dose.

  • 7/29/2019 centraggio in ct

    8/25

  • 7/29/2019 centraggio in ct

    9/25

    Improper positioning cont.Improper positioning cont. Navasivayam et al., (2006) have reported that up

    to 13cm are sometimes exposed unintentionallyin reference radiographs due to inadequatepositioning.

    Increase in radiation dose due to improperplacement of the upper limb in systems usingtube current modulation (up to 45% higher).

    Degradation in image quality usually occurs.

  • 7/29/2019 centraggio in ct

    10/25

  • 7/29/2019 centraggio in ct

    11/25

    What is tube currentWhat is tube current

    modulation??????????modulation?????????? Automatic tube current adjustment along the x-y

    axis (angular) and z-axis (or both) according topatient thickness.

    Philips uses DoseRight-consists of two types of

    modulations ACS and DOM.

    ACS uses a reference radiograph to asses thepatients attenuation. It suggests the lowest mA

    to maintain adequate image quality. mA is set sothat 90% of images will have noise levels equalto or higher than that of a reference image.

  • 7/29/2019 centraggio in ct

    12/25

    Tube current modulationTube current modulation cont.cont.

    D-DOM provides angular modulation.

    Z-DOM modulates mA in the z-axis

  • 7/29/2019 centraggio in ct

    13/25

    Reasons for off centeringReasons for off centering Awareness/attention: lack of education,

    increase in workload etc. Kalra et al.,(2008) demonstrated that in 240 patients,mean mis-positioning averaged 33mm in

    95% of patients (range 5.1-97mm). Adjustment of DFOV: acquisition of

    reference radiographs at different planes.

    Patient related factors: physicalabnormalities, obese individuals etc.

  • 7/29/2019 centraggio in ct

    14/25

    Off centering: effect onOff centering: effect on

    scout datascout data Inappropriate centering effects size of reference

    radiographs. Magnification or reduction of thelatter accompanies mis-operation of DOM.

    The latter leads to erroneous calculations formA settings

  • 7/29/2019 centraggio in ct

    15/25

    Magnification and offMagnification and off

    centeringcentering

  • 7/29/2019 centraggio in ct

    16/25

    Effects of off centeringEffects of off centering CT uses a bow tie filter to shape X-

    ray beam to the body, i.e. restrictphotons at the periphery (areas withless attenuation) and allow morephotons to pass trough the centralportions i.e. those with greater

    attenuation and contribution toimage noise.

  • 7/29/2019 centraggio in ct

    17/25

  • 7/29/2019 centraggio in ct

    18/25

    Off centering and bow tieOff centering and bow tie

    filtersfilters Shaping the radiation beam allows

    for:1)A reduction in scattered radiation

    from peripheral object zones.2)Reduce patient dose.

    3) Improve image quality.

  • 7/29/2019 centraggio in ct

    19/25

    Effects of off centeringEffects of off centering cont.cont.

    BF presumes that the centre of the

    body is at the isocenter. If thepatient is off centre, portions of thebody other than the peripheral

    portions receive fewer photon.

    Increase in image noise and

    radiation dose to peripheries.

  • 7/29/2019 centraggio in ct

    20/25

    With correct centering, more photons(X+++) traverse the thicker central

    parts, resulting in adequate imagenoise and radiation dose.

  • 7/29/2019 centraggio in ct

    21/25

    More photon (X+++) passing throughperipheral parts, increase noise centrally

    and dose peripherally.

  • 7/29/2019 centraggio in ct

    22/25

    Effects of off centeringEffects of off centering cont.cont. Phantom studies have shown a 50%

    increase in surface and peripheral doses(12%-18%@30mm.41%-50%@60mm).

    Actual patients: 2%-30% (older study). Organs mostly at risk: breast, eye lens,

    thyroid. Such phenomena are morecommon when using lower mA, forexample when imaging children.

    Such phenomena are more common whenusing lower mA, for example whenimaging children.

  • 7/29/2019 centraggio in ct

    23/25

    Off centering and imageOff centering and image

    qualityquality

    Difference in image noise @30mm (left)and 60mm (right).

  • 7/29/2019 centraggio in ct

    24/25

    Effect on image qualityEffect on image quality

  • 7/29/2019 centraggio in ct

    25/25

    ConclusionConclusion Radiographers must pay attention

    whilst position/centering. Patients arms should elevated on

    top of the head. When not possible

    improvised arrangements should bemade.

    Radiographers are directly

    responsible for patient doseso thinkbefore you act!